Objective: We tested the suitability of the GORE HELEX septal occluder for closure of membranous ventricular septal defect in three Yucatan minipigs that had naturally-occurring membranous ventricular septal defects.

Methods: The animals were studied in the Gore Science Center Laboratory in Flagstaff, Arizona. In all of the device implantations, seating of the right disk along the right ventricular septum was problematic with resultant entanglement of one device in the chordae tendinae during attempted removal, implantation of one device onto the tricuspid valve papillary muscle and unlocking of one device occurring after implantation.

Conclusions: A new occluder design appears to be needed to minimize the risk of disk entanglement with cardiac structures, particularly within the right ventricle. These studies did establish that the Yucatan mini swine are an acceptable and appropriate model for evaluating occluders intended for closure of membranous ventricular septal defects.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.22601DOI Listing

Publication Analysis

Top Keywords

membranous ventricular
12
ventricular septal
12
gore helex
8
septal defect
8
closure membranous
8
septal
5
failure gore
4
helex atrial
4
atrial septal
4
defect occluder
4

Similar Publications

We report a rare complication of left ventricular giant thrombus in a patient with fulminant myocarditis after venoarterial extracorporeal membrane oxygenation therapy. This case report offers simple anticoagulant treatment experiences to eliminate significant LV thrombosis in patient undergoing extracorporeal membrane oxygenation, so that patients do not need surgery.

View Article and Find Full Text PDF

Background And Objective: It has been believed that polymorphic ventricular tachycardia (VT) such as torsades de pointes (TdP) seen in patients with long QT syndromes is triggered by creating early afterdepolarization (EAD)-mediated triggered activity (TA). Although the mechanisms creating the TA have been studied intensively, characteristics of the arrhythmogenic (torsadogenic) substrates that link EAD developments to TA formation are still not well understood.

Methods: Computer simulations of excitation propagation in a homogenous two-dimensional ventricular tissue with an anisotropic conduction property were performed to characterize torsadogenic substrates that potentially form TA.

View Article and Find Full Text PDF

Cardiac acetylcholinesterase and butyrylcholinesterase have distinct localization and function.

Am J Physiol Heart Circ Physiol

January 2025

Comenius University Bratislava, Faculty of Pharmacy, Department of Pharmacology and Toxicology, Bratislava, Slovakia.

Cholinesterase (ChE) inhibitors are under consideration to be used in the treatment of cardiovascular pathologies. A prerequisite to advancing ChE inhibitors into the clinic is their thorough characterization in the heart. The aim here was to provide a detailed analysis of cardiac ChE to understand their molecular composition, localization, and physiological functions.

View Article and Find Full Text PDF

Unlabelled: Heparin-induced thrombocytopenia (HIT) is an immune-mediated disease with severe thromboembolic complications. HIT during percutaneous coronary intervention (PCI) can be fatal without prompt treatment. We report an unusual case of HIT observed during PCI for acute coronary syndrome (ACS).

View Article and Find Full Text PDF

In every heartbeat, cardiac muscle cells perform excitation-Ca signaling-contraction (EC) coupling to pump blood against the vascular resistance. Cardiomyocytes can sense the mechanical load and activate mechano-chemo-transduction (MCT) mechanism, which provides feedback regulation of EC coupling. MCT feedback is important for the heart to upregulate contraction in response to increased load to maintain cardiac output.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!