An experimental evaluation of a new craniofacial implant using the rabbit tibia model: Part II. Biomechanical findings.

Otol Neurotol

Department of Biomaterials, Institute for Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.

Published: July 2010

Hypothesis: This experimental study was undertaken to test the hypothesis that a new design of a craniofacial implant is more stable and shows faster integration than the currently used implant.

Materials And Methods: A total of 60 newly designed Cochlear Baha titanium implants (test) and 60 of the current design (control) were placed in the tibiae of 30 adult loop-eared rabbits. The animals were euthanized after 5, 14, or 28 days. The stability of the implants was measured with Implant Stability Quotient measurements and removal torque (RTQ) tests. The Student's t test for paired statistics were used for comparisons and a difference considered if p < 0.05.

Results: The test implants showed a significantly higher RTQ at all time points and a steeper increase with time than the controls. The RTQ values were 19.0 (standard deviation [SD], 9.1) and 9.8 (SD, 5.6) Ncm for test and control implants, respectively, after 5 days; 27.6 (SD, 9.6) and 11.2 (SD, 3.9) Ncm after 14 days; and 43.3 (SD, 9.6) and 20.7 (SD, 14.1) Ncm after 28 days. Higher Implant Stability Quotient values were measured for the test implant at all time points. Calculations of shear stress to compensate for implant diameter revealed significantly higher values for test than for control implants after 14 and 28 days.

Conclusion: It is concluded that the new craniofacial implant was more stable and showed a faster integration than the currently used implant. The data from the present study suggest that the new implant may be suitable for early loading protocols.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0b013e3181de4b79DOI Listing

Publication Analysis

Top Keywords

craniofacial implant
12
implant
9
implant stable
8
stable faster
8
faster integration
8
integration currently
8
implant stability
8
stability quotient
8
time points
8
test control
8

Similar Publications

Application of a semi-active robotic system for implant placement in atrophic posterior maxilla: a retrospective case series.

J Dent

January 2025

Key Laboratory of Shaanxi Province for Craniofacial Precision Medicine Research, College of Stomatology, Xi'an Jiaotong University; Clinical Research Center of Shaanxi Province for Dental and Maxillofacial Disease, College of Stomatology, Xi'an Jiaotong University; Department of Implant Dentistry, College of Stomatology, Xi'an Jiaotong University. Electronic address:

Objective: The study aimed to evaluate the accuracy and safety of the semi-active robotic system for implant placement in atrophic posterior maxilla.

Methods: Patients underwent robot-assisted implant placement in atrophic posterior maxilla were identified and included. Cone-beam computed tomography (CBCT) was performed before surgery.

View Article and Find Full Text PDF

 Minimally invasive Ponto surgery (MIPS) enables the installation of percutaneous bone-anchored hearing implants (BAHIs) with a drill guide through a hole punch incision. Despite being well established for adults, there is a lack of studies in the literature regarding its use in pediatric patients.  The aim of the present study was to investigate the hearing performance and soft-tissue outcomes of the use of MIPS under local anesthesia in children with unilateral craniofacial malformation (UCM).

View Article and Find Full Text PDF

Development of Prevascularized Synthetic Block Graft for Maxillofacial Reconstruction.

J Funct Biomater

January 2025

Center for Oral, Clinical and Translational Sciences, Faculty of Dentistry, Oral & Craniofacial Sciences, King's College London, London SE1 9RT, UK.

Cranio-maxillofacial bone reconstruction, especially for large defects, remains challenging. Synthetic biomimetic materials are emerging as alternatives to autogenous grafts. Tissue engineering aims to create natural tissue-mimicking materials, with calcium phosphate-based scaffolds showing promise for bone regeneration applications.

View Article and Find Full Text PDF

Polyetheretherketone (PEEK) is a high-performance polymer material for developing varying orthopedic, spine, cranial, maxillofacial, and dental implants. Despite their commendable mechanical properties and biocompatibility, the major limitation of PEEK implants is their low affinity to osseointegrate with the neighboring bone. Over the last two decades, several efforts have been made to incorporate bioactive components such as bioceramic particles in PEEK to enhance its osseointegration capacity.

View Article and Find Full Text PDF

Cardiovascular disease (CVD) is the leading cause of death worldwide and has been confirmed to be associated with a common oral bacterial infection-chronic apical periodontitis (CAP). However, the detailed mechanisms remain controversial. CAP can potentially alter systemic inflammation, lipid metabolism, and gut microbiota, all of which contribute to the progression of the aortic inflammatory response.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!