Cardiopulmonary bypass (CPB) is widely used for cardiac surgery by virtue of its proven safety over the course of its use during the past half century. Even though perfusion is safer, incidents still occur. During the repair of a ventricular-septal defect in an 11-month-old infant, we experienced a critical incident related to the potential hazardous effect of volatile anesthetics on the polycarbonate connector of extra-corporeal circuit. The damage to the polycarbonate connector had occurred after spillage of isoflurane during the filling of the vaporizer, causing it to crack and leak. The incident was managed by replacement of the cracked connector during a temporary circulatory arrest. The patient was hypothermic and the time off bypass was less than 1.5 min. There were no neurologic sequelae, the patient made an uneventful recovery. In conclusion, the spillage of volatile anesthetics can cause cracks in the polycarbonate connector of the extra-corporeal circuit, leading to potentially interruption of CPB.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2872842PMC
http://dx.doi.org/10.4097/kjae.2010.58.3.304DOI Listing

Publication Analysis

Top Keywords

polycarbonate connector
16
connector extra-corporeal
12
extra-corporeal circuit
12
cracks polycarbonate
8
volatile anesthetics
8
connector
5
isoflurane cracks
4
polycarbonate
4
circuit case
4
case report-
4

Similar Publications

Platelet Adhesion and Activation in an ECMO Thrombosis-on-a-Chip Model.

Adv Sci (Weinh)

August 2024

School of Medical Sciences, Faculty of Medicine and Health, The University of Sydney, Sydney, NSW, 2006, Australia.

Use of extracorporeal membrane oxygenation (ECMO) for cardiorespiratory failure remains complicated by blood clot formation (thrombosis), triggered by biomaterial surfaces and flow conditions. Thrombosis may result in ECMO circuit changes, cause red blood cell hemolysis, and thromboembolic events. Medical device thrombosis is potentiated by the interplay between biomaterial properties, hemodynamic flow conditions and patient pathology, however, the contribution and importance of these factors are poorly understood because many in vitro models lack the capability to customize material and flow conditions to investigate thrombosis under clinically relevant medical device conditions.

View Article and Find Full Text PDF

Visual Sensor for Sterilization of Polymer Fixtures Using Embedded Mesoporous Silicon Photonic Crystals.

ACS Sens

January 2018

Sienna Biopharmaceuticals Inc. , 30699 Russell Ranch Road, Suite 140, Westlake Village, California 91362, United States.

A porous photonic crystal is integrated with a plastic medical fixture (IV connector hub) to provide a visual colorimetric sensor to indicate the presence or absence of alcohol used to sterilize the fixture. The photonic crystal is prepared in porous silicon (pSi) by electrochemical anodization of single crystal silicon, and the porosity and the stop band of the material is engineered such that the integrated device visibly changes color (green to red or blue to green) when infiltrated with alcohol. Two types of self-reporting devices are prepared and their performance compared: the first type involves heat-assisted fusion of a freestanding pSi photonic crystal to the connector end of a preformed polycarbonate hub, forming a composite where the unfilled portion of the pSi film acts as the sensor; the second involves generation of an all-polymer replica of the pSi photonic crystal by complete thermal infiltration of the pSi film and subsequent chemical dissolution of the pSi portion.

View Article and Find Full Text PDF

Use of Volatile Anesthetic Agent in Extracorporeal Circuit as a Cause of Break in Polycarbonate Connector-Lessons Learnt.

J Extra Corpor Technol

September 2017

Department of Cardiothoracic and Vascular Surgery, Sri Sathya Sai Institute of Higher Medical Sciences, Prasanthigram, Puttaparthy, Anantapur District, Andhra Pradesh, India.

Mishaps, near misses, and lethal incidents are known to occur during cardiopulmonary bypass. We share one such rare case of break in polycarbonate connector because of the use of isoflurane in extracorporeal circuit and its successful management.

View Article and Find Full Text PDF

Piezoelectric Micro- and Nanostructured Fibers Fabricated from Thermoplastic Nanocomposites Using a Fiber Drawing Technique: Comparative Study and Potential Applications.

ACS Nano

February 2017

Génie Métallurgique and ‡Génie Physique, École Polytechnique de Montréal , Montréal, Québec H3T 1J4, Canada.

We report an all-polymer flexible piezoelectric fiber that uses both judiciously chosen geometry and advanced materials in order to enhance fiber piezoelectric response. The microstructured/nanostructured fiber features a soft hollow polycarbonate core surrounded by a spiral multilayer cladding consisting of alternating layers of piezoelectric nanocomposites (polyvinylidene enhanced with BaTiO, PZT, or CNT) and conductive polymer (carbon-filled polyethylene). The conductive polymer layers serve as two electrodes, and they also form two spatially offset electric connectors on the fiber surface designed for the ease of connectorization.

View Article and Find Full Text PDF

Cardiopulmonary bypass (CPB) is widely used for cardiac surgery by virtue of its proven safety over the course of its use during the past half century. Even though perfusion is safer, incidents still occur. During the repair of a ventricular-septal defect in an 11-month-old infant, we experienced a critical incident related to the potential hazardous effect of volatile anesthetics on the polycarbonate connector of extra-corporeal circuit.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!