Objective: To compare breath-hold 1H-magnetic resonance spectroscopy (1H-MRS) with respiratory-gated 1H-MRS and computed tomography (CT) for quantification of hepatic lipid content.

Methods: Twenty-three premenopausal women underwent breath-hold point-resolved single-voxel 1H-MRS of the liver followed by respiratory-gated 1H-MRS at 3 Tesla and CT slice through the liver. Interscan variability for 1H-MRS was assessed in 6 volunteers. Pearson correlation coefficients, Bland-Altman 95% limit of agreement, and concordance correlation coefficients were calculated.

Results: There was a strong correlation between breath-hold and respiratory-gated 1H-MRS (r = 0.94, P < 0.0001; concordance correlation coefficient, 0.75). Using Bland-Altman analysis, all but 2 data points were within the limits of agreement. Both 1H-MRS techniques had low interscan variability. There was an inverse correlation of both 1H-MRS techniques with CT attenuation values of the liver.

Conclusions: Breath-hold 1H-MRS is a reliable method to measure hepatic lipid content at 3 Tesla. Breath-hold 1H-MRS of the liver provides data that closely correlates with that obtained from longer-duration respiratory-gated technique.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2877282PMC
http://dx.doi.org/10.1097/RCT.0b013e3181cefb89DOI Listing

Publication Analysis

Top Keywords

respiratory-gated 1h-mrs
12
1h-mrs
10
breath-hold 1h-magnetic
8
1h-magnetic resonance
8
resonance spectroscopy
8
hepatic lipid
8
1h-mrs liver
8
interscan variability
8
correlation coefficients
8
concordance correlation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!