Background: Capsule endoscopy that could actively move and approach a specific site might be more valuable for the diagnosis or treatment of GI diseases.

Objective: We tested the performance of active locomotion of a novel wired capsule endoscope with a paddling-based locomotion mechanism, using 3 models: a silicone tube, an extracted porcine colon, and a living pig.

Design: In vitro, ex vivo, and in vivo experiments in a pig model.

Setting: Study in an animal laboratory.

Interventions: For the in vitro test, the locomotive capsule was controlled to actively move from one side of a silicone tube to the other by a controller-operated automatic traveling program. The velocity was calculated by following a video recording. We performed ex vivo tests by using an extracted porcine colon in the same manner we performed the in vitro test. In in vivo experiments, the capsule was inserted into the rectum of a living pig under anesthesia, and was controlled to move automatically forward. After 8 consecutive trials, the velocity was calculated.

Main Outcome Measurements: Elapsed time, velocity, and mucosal damage.

Results: The locomotive capsule showed stable and active movement inside the lumen both in vitro and ex vivo. The velocity was 60 cm/min in the silicone tube, and 36.8 and 37.5 cm/min in the extracted porcine colon. In the in vivo experiments, the capsule stably moved forward inside the colon of a living pig without any serious complications. The mean velocity was 17 cm/min over 40 cm length. We noted pinpoint erythematous mucosal injuries in the colon.

Limitation: Porcine model experiments, wired capsule endoscope.

Conclusions: The novel paddling-based locomotive capsule endoscope performed fast and stable movement in a living pig colon with consistent velocity. Further investigation is necessary for practical use in humans.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.gie.2009.12.058DOI Listing

Publication Analysis

Top Keywords

capsule endoscope
12
vitro vivo
12
silicone tube
12
extracted porcine
12
porcine colon
12
vivo experiments
12
locomotive capsule
12
living pig
12
capsule
9
active locomotion
8

Similar Publications

Reducing reading time and assessing disease in capsule endoscopy videos: A deep learning approach.

Int J Med Inform

January 2025

University of Coimbra, Faculty of Medicine, Coimbra, Portugal; Department of Gastroenterology, Centro Hospitalar e Universitário de Coimbra, Coimbra, Portugal. Electronic address:

Background: The wireless capsule endoscope (CE) is a valuable diagnostic tool in gastroenterology, offering a safe and minimally invasive visualization of the gastrointestinal tract. One of the few drawbacks identified by the gastroenterology community is the time-consuming task of analyzing CE videos.

Objectives: This article investigates the feasibility of a computer-aided diagnostic method to speed up CE video analysis.

View Article and Find Full Text PDF

Background: An all-inside endoscopic flexor hallucis longus (FHL) tendon transfer is indicated for the treatment of chronic, full-thickness Achilles tendon defects. The aim of this procedure is to restore function of the gastrocnemius-soleus complex while avoiding the wound complications associated with open procedures.

Description: This procedure can be performed through 2 endoscopic portals, a posteromedial portal (the working portal) and a posterolateral portal (the visualization portal).

View Article and Find Full Text PDF

Background: Craniopharyngiomas are epithelial tumors derived from the remnants of the Rathke pouch, while Rathke cleft cysts (RCC) are benign cystic lesions originating from the Rathke pouch itself [1]. Rathke cleft cysts comprise 10-15% of the hypophyseal tumors, while craniopharyngiomas are relatively rare, comprising only 2-5% of intracranial tumors [2]. Both located in the sellar and parasellar regions and share clinical symptoms including headache, visual disturbances, and endocrine dysfunction [3].

View Article and Find Full Text PDF

Wide FOV metalens for near-infrared capsule endoscopy: advancing compact medical imaging.

Nanophotonics

November 2024

National Key Laboratory on Near-Surface Detection, School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China.

This study presents the design, fabrication, and characterization of a wide field-of-view (FOV) metalens optimized for capsule endoscopy. The metalens achieved a 165° FOV with a high modulation transfer function (MTF) of 300 lines per millimeter (lp/mm) across the entire FOV, operating in the near-infrared (NIR) narrow-bandpass imaging at 940 nm. The performance of the metalens-based system is evaluated using two bandwidths, 12 nm and 32 nm, showing MTF values of 0.

View Article and Find Full Text PDF

The living human inner ear is challenging to study because it is encased within dense otic capsule bone that limits access to biological tissue. Traditional temporal bone histopathology methods rely on lengthy, expensive decalcification protocols that take 9-10 months and reduce the types of tissue analysis possible due to RNA degradation. There is a critical need to develop methods to access fresh human inner ear tissue to better understand otologic diseases, such as Ménière's disease, at the cellular and molecular level.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!