We have identified a new, nanosecond pulsed electric field (nsPEF) therapy capable of eliminating murine melanomas located in the skin with a single treatment. When these optimized parameters are used, nsPEFs initiate apoptosis without hyperthermia. We have developed new suction electrodes that are compatible with human skin and have applied them to a xenograft nude mouse melanoma model system to identify the optimal field strength, pulse frequency and pulse number for the treatment of murine melanomas. A single treatment using the optimal pulse parameters (2,000 pulses, 100 ns in duration, 30 kV/cm in amplitude at a pulse frequency of 5-7 pulses/sec) eliminated all 17 melanomas treated with those parameters in 4 mice. This was the highest pulse frequency that we could use without raising the treated skin tumor temperature above 40 degrees C. We also demonstrate that the effects of nsPEF therapy are highly localized to only cells located between electrodes and results in very little scarring of the nsPEF-treated skin.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2919636PMC
http://dx.doi.org/10.1002/ijc.25364DOI Listing

Publication Analysis

Top Keywords

single treatment
12
pulse frequency
12
nanosecond pulsed
8
pulsed electric
8
electric field
8
nspef therapy
8
murine melanomas
8
pulse
5
optimized nanosecond
4
field therapy
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?