[Feedback control of depth of anesthesia during propofol administration. Bispectral index as the controlled variable].

Anaesthesist

Klinik und Poliklinik für Anästhesiologie und Intensivtherapie, Universität Rostock, Schillingallee 35, 18055 Rostock, Deutschland.

Published: July 2010

Background: The objective of this study was to evaluate the performance of a new system for closed-loop control of propofol administration using the bispectral index (BIS) under total intravenous anesthesia in the index values of middle-to-deep depth of anaesthesia.

Methods: In this study 20 adult patients anesthetized with propofol and remifentanil were investigated. The propofol infusion was carried out using a fuzzy-PD+I controller with a target BIS value of 40.

Results: Closed-loop control was able to provide maintenance of anesthesia and adequate operating conditions for all patients. The following quality control criteria were calculated: median performance error (MDPE; 0.16%, SD +/-1.4%), median absolute performance error (MDAPE; 6.9%, SD +/-2.8%) and wobble (6.8%, SD +/-2.5%).

Conclusion: The present study showed the clinical feasibility of the controller compared to existing devices regarding a high level of quality criteria of a model with an implemented fuzzy-PD+I structure controlling depth of hypnosis.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00101-010-1731-9DOI Listing

Publication Analysis

Top Keywords

propofol administration
8
administration bispectral
8
closed-loop control
8
performance error
8
[feedback control
4
control depth
4
depth anesthesia
4
propofol
4
anesthesia propofol
4
bispectral controlled
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!