Background: We developed a new disposable anaerobic culture system, namely, the Quick anaero-system, for easy culturing of obligate anaerobes.

Methods: Our system consists of 3 components: 1) new disposable anaerobic gas pack, 2) disposable culture-envelope and sealer, and 3) reusable stainless plate rack with mesh containing 10 g of palladium catalyst pellets. To evaluate the efficiency of our system, we used 12 anaerobic bacteria. We prepared 2 sets of ten-fold serial dilutions of the 12 anaerobes, and inoculated these samples on Luria-Bertani (LB) broth and LB blood agar plate (LB-BAP) (BD Diagnostic Systems, USA). Each set was incubated in the Quick anaero-system (DAS Tech, Korea) and BBL GasPak jar with BD GasPak EZ Anaerobe Container System (BD Diagnostic Systems) at 35-37 degrees C for 48 hr. The minimal inoculum size showing visible growth of 12 anaerobes when incubated in both the systems was compared.

Results: The minimal inoculum size showing visible growth for 2 out of the 12 anaerobes in the LB broth and 9 out of the 12 anaerobes on LB-BAP was lower for the Quick anaero-system than in the BD GasPak EZ Anaerobe Container System. The mean time (+/-SD) required to achieve absolute anaerobic conditions of the Quick anaero-system was 17 min and 56 sec (+/-3 min and 25 sec).

Conclusions: The Quick anaero-system is a simple and effective method of culturing obligate anaerobes, and its performance is superior to that of the BD GasPak EZ Anaerobe Container System.

Download full-text PDF

Source
http://dx.doi.org/10.3343/kjlm.2010.30.2.133DOI Listing

Publication Analysis

Top Keywords

quick anaero-system
20
disposable anaerobic
12
gaspak anaerobe
12
anaerobe container
12
container system
12
anaerobic culture
8
culture system
8
culturing obligate
8
diagnostic systems
8
minimal inoculum
8

Similar Publications

Background: We developed a new disposable anaerobic culture system, namely, the Quick anaero-system, for easy culturing of obligate anaerobes.

Methods: Our system consists of 3 components: 1) new disposable anaerobic gas pack, 2) disposable culture-envelope and sealer, and 3) reusable stainless plate rack with mesh containing 10 g of palladium catalyst pellets. To evaluate the efficiency of our system, we used 12 anaerobic bacteria.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!