Purpose: (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer.

Methods: A trained technician used the ICare rebound tonometer to measure the IOP of the right eye of 100 glaucoma patients. The technician then instructed each patient on use of the ICare tonometer. Each patient then measured his/her own pressure using the ICare tonometer. Finally, a different technician, who was masked to both of the earlier readings, measured IOP by Goldmann applanation tonometry. Thirty patients repeated the ICare measurement 3 times (once every 5 min) 20 minutes after the initial IOP measurement.

Results: Of the 100 patients, 82 of patient ICare and the technician ICare readings were within 3 mm Hg of each other, and 75 of the patient ICare and Goldmann applanation tonometry measurements were within 3 mm Hg of each other. Intraclass correlations between self-administered ICare measurements 1 and 2, 1 and 3, and 2 and 3 were 0.69, 0.71 and 0.81, respectively.

Conclusion: In this study, the ICare rebound tonometer was accurate and reliable in the hands of patients. Patients can easily learn to self-administer this test, possibly allowing for home monitoring of IOP.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IJG.0b013e3181d78811DOI Listing

Publication Analysis

Top Keywords

icare rebound
20
rebound tonometer
16
icare
12
patient icare
12
intraocular pressure
8
icare measurements
8
trained technician
8
technician icare
8
icare tonometer
8
goldmann applanation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!