A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Biodisk: a new device for closure of patent foramen ovale: a feasibility study in swine. | LitMetric

Purpose: To evaluate the feasibility, effectiveness, and safety of a porcine small intestinal submucosa (SIS)-covered Biodisk (BD) for the closure of patent foramen ovale (PFO) in swine.

Methods: Twelve piglets (9-30 kg) with PFO ranging in size from 6 to 12 mm were used for the in vivo testing. The BD device consisted of two basic nitinol wire components covered with platinum coil, a flexible SIS-covered ring, and an anchor. The BD was advanced through an 8-Fr sheath from the femoral vein. Nine acute animals were used to test the BD for deployment, stability, immediate shunt closure, and device repositioning before or after its detachment. To assess retrievability, four devices were deployed and intentionally embolized into the RA (n = 2) and LA (n = 2). The effectiveness of the device was evaluated by angiocardiography. EKG was recorded before and after PFO closure for 3 hr. From the 12 animals, nine were acute and three were followed; one for 6 weeks, one for 12 weeks, and one for 16 weeks.

Results: Successful device implantation was achieved in all animals with no shunting of contrast media observed during follow-up in. One animal needed to have device repositioned for complete PFO occlusion because of suboptimal placement at the first attempt. The device was easily placed and retrieved before detachment in all nine animals in the acute study. None of the BDs spontaneously embolized during release or on follow-up. EKG did not demonstrate arrhythmias during or after treatment. Four intentionally embolized BDs were easily retrieved with an Amplatz goose neck snare. Macroscopic and histologic evaluation of the three long-term animals showed that devices were well incorporated in the atrial septum with complete shunt closure. The SIS showed progressive remodeling with the host cells. There was also progressive endothelization of the BD device.

Conclusion: The BD device deployment is feasible, safe, and effective. Long-term studies are needed to evaluate its long-term effectiveness.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.22429DOI Listing

Publication Analysis

Top Keywords

closure patent
8
patent foramen
8
foramen ovale
8
shunt closure
8
intentionally embolized
8
animals acute
8
easily retrieved
8
device
7
closure
5
animals
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!