Pulse conductance and flow-induced hemolysis during pulsatile cardiopulmonary bypass.

Artif Organs

Department of Cardiothoracic Surgery and Cardiovascular Research Institute Maastricht, Maastricht University Medical Center, Maastricht, The Netherlands.

Published: April 2010

In this study, the hypothesis was tested that a low-resistant, high-compliant oxygenator provides better pulse conductance and less hemolysis than a high-resistant, low-compliant oxygenator during pulsatile cardiopulmonary bypass. Forty adults undergoing coronary artery bypass surgery were randomly divided into two groups using either an oxygenator with a relatively low hydraulic resistance (Quadrox BE-HMO 2000, Maquet Cardiopulmonary AG, Hirrlingen, Germany) or with a relatively high hydraulic resistance (Capiox SX18, Terumo Cardiovascular Systems, Tokyo, Japan). The phase shift between the flow signals measured at the inlet and outlet of the oxygenator was used to assess compliance. Pulse conductance in terms of pressure attenuation was calculated by dividing the outlet pulse pressure of the oxygenator by the inlet pulse pressure. A normalized index was used to assess hemolysis. The phase shifts in time of the flow pulses were 36 +/- 6 ms in the low-resistant (high-compliant) oxygenator, and 14 +/- 2 ms in the high-resistant (low-compliant) oxygenator group (P < 0.001). The low-resistant, high-compliant oxygenator provided 27% better pulse conductance compared with the high-resistant, low-compliant oxygenator (0.84 +/- 0.02 and 0.66 +/- 0.01, respectively, P < 0.001). Inlet pulse pressures were significantly higher (29%) in the high-resistant, low-compliant (Capiox) group than in the low-resistant, high-compliant (Quadrox) group (838 +/- 38 mm Hg and 648 +/- 25 mm Hg respectively, P < 0.001), but no significant difference in hemolysis was found. A low-resistant, high-compliant oxygenator provides better pulse conduction than a high-resistant, low-compliant oxygenator. However, the study data could not confirm the association of high pressures with increased hemolysis.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1525-1594.2009.00853.xDOI Listing

Publication Analysis

Top Keywords

low-resistant high-compliant
20
high-resistant low-compliant
20
pulse conductance
16
high-compliant oxygenator
16
low-compliant oxygenator
16
better pulse
12
oxygenator
11
pulse
8
pulsatile cardiopulmonary
8
cardiopulmonary bypass
8

Similar Publications

Pulse conductance and flow-induced hemolysis during pulsatile cardiopulmonary bypass.

Artif Organs

April 2010

Department of Cardiothoracic Surgery and Cardiovascular Research Institute Maastricht, Maastricht University Medical Center, Maastricht, The Netherlands.

In this study, the hypothesis was tested that a low-resistant, high-compliant oxygenator provides better pulse conductance and less hemolysis than a high-resistant, low-compliant oxygenator during pulsatile cardiopulmonary bypass. Forty adults undergoing coronary artery bypass surgery were randomly divided into two groups using either an oxygenator with a relatively low hydraulic resistance (Quadrox BE-HMO 2000, Maquet Cardiopulmonary AG, Hirrlingen, Germany) or with a relatively high hydraulic resistance (Capiox SX18, Terumo Cardiovascular Systems, Tokyo, Japan). The phase shift between the flow signals measured at the inlet and outlet of the oxygenator was used to assess compliance.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!