Introduction: Difficult airway management in children is challenging. One alternative device to the gold standard of direct laryngoscopy is the STORZ Bonfils fiberscope (Karl Storz Endoscopy, Tuttlingen, Germany), a rigid fiberoptic stylette-like scope with a curved tip. Although results in adults have been encouraging, reports regarding its use in children have been conflicting. We compared the effectiveness of a standard laryngoscope to the Bonfils fiberscope in a simulated difficult infant airway.

Methods: Ten pediatric anesthesiologists were recruited for this study and asked to perform three sets of tasks. For the first task, each participant intubated an unaltered manikin (SimBaby (TM), Laerdal, Puchheim, Germany) five times using a styletted 3.5 endotracheal tube (ETT) and a Miller 1 blade (group DL-Normal). For the second task, a difficult airway configuration simulating a Cormack-Lehane grade 3B view was created by fixing a Miller-1 blade into position in the manikin using a laboratory stand. Each participant then intubated the manikin five times with a styletted 3.5 ETT using conventional technique but without touching the laryngoscope (group DL-Difficult). In the third task, the manikin was kept in the same difficult airway configuration, and each participant intubated the manikin five times using a 3.5-mm ETT mounted on the Bonfils fiberscope as an adjunct to direct laryngoscopy with the Miller-1 blade (group BF-Difficult). Primary outcomes were time to intubate and success rate.

Results: A total of 150 intubations were performed. Correct ETT placement was achieved in 100% of attempts in group DL-Normal, 90% of attempts in group DL-Difficult and 98% of attempts in BF-Difficult. Time to intubate averaged 14 s (interquartile range 12-16) in group DL-Normal; 12 s (10-15) in group DL-Difficult; and 11 s (10-18) in group BF-Difficult. The percentage of glottic opening seen (POGO score) was 70% (70-80) in group DL-Normal; 0% (0-0) in group DL-Difficult; and 100% (100-100) in group BF-Difficult.

Discussion: The Bonfils fiberscope-assisted laryngoscopy was easier to use and provided a better view of the larynx than simple direct laryngoscopy in the simulated difficult pediatric airway, but intubation success rate and time to intubate were not improved. Further studies of the Bonfils fibrescope as a pediatric airway adjunct are needed.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1460-9592.2010.03298.xDOI Listing

Publication Analysis

Top Keywords

direct laryngoscopy
16
group dl-normal
16
group dl-difficult
16
simulated difficult
12
difficult airway
12
bonfils fiberscope
12
participant intubated
12
time intubate
12
group
11
bonfils fiberscope-assisted
8

Similar Publications

Background: Securing a stable airway is a critical component in neonatal resuscitation. Compared to direct laryngoscopy, video laryngoscopy provides improved visualization of the glottis, potentially enhancing the success rate of intubation. This systematic review and meta-analysis were conducted to assess and compare the efficacy and safety of video laryngoscopy versus direct laryngoscopy in neonatal intubation.

View Article and Find Full Text PDF

Background: Benign laryngeal lesions, characterized by non-cancerous growths in the larynx, significantly impact voice quality and respiratory function. These lesions, which include vocal cord polyps, nodules, papillomas, and cysts, often result from factors such as vocal abuse, viral infections, and chronic inflammation. While studies on benign laryngeal lesions are well-documented globally, data specific to Northern Nigeria remains sparse.

View Article and Find Full Text PDF

Objectives: Injection laryngoplasty (IL) is commonly performed for glottic insufficiency and has historically been performed under general anesthesia via direct laryngoscopy (DL), with an increasing number of procedures being performed in the office setting via flexible laryngoscopy (FL). Existing literature regarding complications of IL primarily addresses immediate peri-procedural complications and adverse reactions to the injection material. This is the first study utilizing a large multi-institutional database comparing complications of IL performed via DL versus FL.

View Article and Find Full Text PDF

Background: The Neonatal Resuscitation Program recommends direct laryngoscopy (DL) as the primary method for neonatal intubation. Video laryngoscopy (VL) is suggested as an option, particularly for training novice operators or for intubating infants with difficult airways. The programme outlines specific steps for intubation, including managing the external environment and techniques for visualising key anatomical landmarks.

View Article and Find Full Text PDF

Background: Airway management is a critical component of the care of patients experiencing cardiac arrest, but data from randomized trials on the use of video vs direct laryngoscopy for intubation in the setting of cardiac arrest are limited. Current AHA guidelines recommend placement of an endotracheal tube either during CPR or shortly after return of spontaneous circulation but do not provide guidance around intubation methods, including the choice of laryngoscope.

Research Question: Does use of video laryngoscopy improve the incidence of successful intubation on the first attempt, compared to use of direct laryngoscopy, among adults undergoing tracheal intubation after experiencing cardiac arrest?

Study Design And Methods: This secondary analysis of the Direct versus Video Laryngoscope (DEVICE) trial compared video laryngoscopy versus direct laryngoscopy in the subgroup of patients who were intubated following cardiac arrest.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!