DNA amplification is a fundamental yet costly process used in DNA analysis. This study evaluated half-reaction amplification (12.5, 12, and 13 microL) using the Promega Powerplex 16 Kit with the hope of reducing sample analysis costs by half. A sensitivity study was completed, along with the testing of various blood stain samples including those with low (<0.40 ng) and high DNA concentrations (>3.0 ng), peak height imbalances, and allelic drop-out. Also, 467 samples submitted to the MUFSC laboratory for testing were analyzed. Results indicate that half-reaction amplification produced higher quality profiles than full-reactions. Average peak heights increased by 85%, peak height imbalances improved, and drop-out was eliminated in 75.8% of samples. Only eight of 467 case samples required re-amplification, a success rate of 94% was observed, and the repeat rate decreased significantly. Finally, a DNA input of 0.25-1.0 ng is ideal for half-reaction amplification.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1556-4029.2010.01388.x | DOI Listing |
Forensic Sci Int Genet
September 2019
Institute of Environmental Science and Research Ltd (ESR), Mt Albert Science Centre, Hampstead Road, Private Bag 92021, Auckland, New Zealand. Electronic address:
J Forensic Sci
July 2010
Marshall University Forensic Science Center, DNA Laboratory, 1401 Forensic Science Drive, Huntington, WV, USA.
DNA amplification is a fundamental yet costly process used in DNA analysis. This study evaluated half-reaction amplification (12.5, 12, and 13 microL) using the Promega Powerplex 16 Kit with the hope of reducing sample analysis costs by half.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!