The performance of a four-element Si drift detector for energy-dispersive fluorescence-yield X-ray absorption fine-structure measurements is reported, operating at the National Institute of Standards and Technology beamline X23A2 at the National Synchrotron Light Source. The detector can acquire X-ray absorption fine-structure spectra with a throughput exceeding 4 x 10(5) counts per second per detector element (>1.6 x 10(6) total counts per second summed over all four channels). At this count rate the resolution at 6 keV is approximately 220 eV, which adequately resolves the Mn Kalpha and Kbeta fluorescence lines. Accurate dead-time correction is demonstrated, and it has been incorporated into the ATHENA data analysis program. To maintain counting efficiency and high signal to background, it is suggested that the incoming count rate should not exceed approximately 70% of the maximum throughput.