Aim Of The Study: To investigate if body temperature as measured with a prototype of a non-invasive continuous cerebral temperature sensor using the zero-heat-flow method to reflect the oesophageal temperature (core temperature) during mild therapeutic hypothermia after cardiac arrest.
Methods: In patients over 18 years old with restoration of spontaneous circulation after cardiac arrest, a temperature sensor that uses the zero-heat-flow principle was placed on the forehead during the periods of cooling and re-warming. This temperature was compared to oesophageal temperature as the primary temperature-monitoring site. To assess agreement, we used the Bland-Altman approach and Lin's concordance correlation coefficient.
Results: From September 2008 to April 2009, data from 19 patients were analysed. The median time from restoration of spontaneous circulation until temperature sensor application was 53min (interquartile range, 31; 96). All sensors were removed when a core temperature of 36 degrees C was reached. These measurements were in agreement with oesophageal temperature measurements. No allergic reaction, rash or other irritation occurred on the skin around or under the probes. Bland-Altman results showed a bias of -0.12 degrees C and 95% limits of agreement of -0.59 and +0.36 degrees C. Lin's concordance correlation coefficient was 0.98.
Conclusions: Body temperature measurements using a non-invasive continuous cerebral temperature sensor prototype that uses the zero-heat-flow method accurately reflected oesophageal temperature measurements during mild therapeutic hypothermia in patients with restoration of spontaneous circulation after cardiac arrest.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.resuscitation.2010.03.018 | DOI Listing |