Effect of blower-mister devices on vasoreactivity of coronary artery bypass grafts.

J Thorac Cardiovasc Surg

Division of Cardiology, Department of Internal Medicine II, Medical University of Vienna, Austria.

Published: October 2010

Objective: Potential disadvantages, such as lower graft patency rates, that could lead to the need for more subsequent revascularization after off-pump coronary artery bypass grafting are discussed. One mechanism may involve endothelial damage originating from tools to facilitate a bloodless surgical field during anastomosis. We investigated the effect of the use of a blower-mister device on in vitro endothelial and contractile functions of human internal thoracic arteries and greater saphenous veins.

Methods: The distal ends of both bypass graft types were exposed to the air and water stream of the device for 10 minutes, representing the usual time needed for completion of an anastomosis using the off-pump technique. Vessel segments of both groups were cut into 4-mm rings and stored in a modified Krebs-Henseleit solution. Contraction responses to U44619 and relaxation responses to acetylcholine, substance P, papaverine, and nitroprusside were evaluated.

Results: Forty-eight artery ring segments from 8 patients were studied. Absolute maximum contraction to U44619 was significantly less in rings subjected to the blower-mister device than in controls (internal thoracic artery: 17.17 ± 2.57 mN vs 8.67 ± 4.54 mN, P < .048; greater saphenous vein: 28.33 ± 9.71 mN vs 11.42 ± 7.97 mN, P < .026). Control rings had significantly greater endothelium-dependent relaxation response to acetylcholine (mean difference 29.2% ± 3.4%, P < .001), whereas those subjected to the blower-mister device had reduced responses. Endothelium-independent relaxation to nitroprusside was not significantly different among the groups.

Conclusions: Vessels exposed to the air and water stream of a blower-mister device showed a reduced vasoreactivity. This effect should be studied further, especially if it contributes to lower graft patency rates in off-pump surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jtcvs.2010.02.016DOI Listing

Publication Analysis

Top Keywords

blower-mister device
16
coronary artery
8
artery bypass
8
lower graft
8
graft patency
8
patency rates
8
internal thoracic
8
greater saphenous
8
exposed air
8
air water
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!