Radiation exposure of humans generally results in low doses delivered at low dose rate. Our limited knowledge of the biological effects of low dose radiation is mainly based on data from the atomic bomb Life Span Study (LSS) cohort. However, the total doses and dose rates in the LSS cohort are still higher than most environmental and occupational exposures in humans. Importantly, the dose rate is a critical determinant of health risks stemming from radiation exposure. Understanding the shape of the dose-rate response curve for different biological outcomes is thus crucial for projecting the biological hazard from radiation in different environmental and man-made conditions. A significant barrier to performing low dose-rate studies is the difficulty in creating radiation source configurations compatible with long-term cellular or animal experiments. In this study the design and characterization of a large area, I-based irradiator is described. The irradiator allows continuous long-term exposure of mice at variable dose rates and can be sited in standard animal care facilities. The dose rate is determined by the level of I activity added to a large NaOH-filled rectangular phantom. The desired dose rate is maintained at essentially constant levels by weekly additions of I to compensate for decay. Dosimetry results for long-term animal irradiation at targeted dose rates of 0.00021 and 0.0021 cGy min are presented.