Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom.

Radiother Oncol

Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Tlalpan, Mexico City, Mexico.

Published: August 2010

This work reports the use of both GafChromic EBT film immersed in a water phantom and Monte Carlo (MC) simulations for small photon beam stereotactic radiosurgery dosimetry. Circularly collimated photon beams with diameters in the 4-20 mm range of a dedicated 6 MV linear accelerator (Novalis, BrainLAB, Germany) were used to perform off-axis ratios, tissue maximum ratios and total scatter factors measurements, and MC simulations. GafChromic EBT film data show an excellent agreement with MC results (<2.7%) for all measured quantities.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2010.03.016DOI Listing