A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Accuracy assessment of the Dinamap ProCare 400 in pregnancy and preeclampsia. | LitMetric

Accuracy assessment of the Dinamap ProCare 400 in pregnancy and preeclampsia.

Hypertens Pregnancy

King's College London, Division of Reproduction and Endocrinology, St. Thomas' Hospital, London, United Kingdom.

Published: January 2010

AI Article Synopsis

  • The study evaluated the accuracy of the Dinamap ProCare 400 blood pressure monitor in pregnant women and those with preeclampsia, following the British Hypertension Society guidelines.
  • Ethical approval allowed the recruitment of 45 women, with measurements taken using both the Dinamap and a mercury sphygmomanometer for comparison.
  • The results showed the device received an A/A rating for pregnancy and A/B for preeclampsia, indicating it is generally reliable, though users should exercise caution when interpreting readings.

Article Abstract

Objective: To assess the accuracy of the Dinamap ProCare 400 in pregnancy and preeclampsia, according to the British Hypertension Society protocol.

Methods: Ethical approval was obtained. Forty-five women were recruited at two large teaching hospitals. Nine sequential same arm measurements were taken from each woman alternating between the Dinamap and mercury sphygmomanometry.

Results: The device achieved an A/A grade (pregnancy) and an A/B grade (preeclampsia). Mean difference (SD) were 1.1(5.8)/0.1(5.7) mmHg (pregnancy) and -0.5(5.4)/0.9(9.5) mmHg (preeclampsia) for systolic/diastolic.

Conclusion: The Dinamap ProCare 400 is recommended for use in pregnancy and has good accuracy in preeclampsia, although caution is still advised.

Download full-text PDF

Source
http://dx.doi.org/10.3109/10641950902968650DOI Listing

Publication Analysis

Top Keywords

dinamap procare
12
procare 400
12
400 pregnancy
8
pregnancy preeclampsia
8
pregnancy
5
preeclampsia
5
accuracy assessment
4
dinamap
4
assessment dinamap
4
preeclampsia objective
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!