Objective: Image quality of chest and abdominal computed tomographic (CT) scans was evaluated at different doses to assess the lowest value of x-ray dose at which the image quality was not being affected.

Materials And Methods: Using multislice CT (MSCT), 29 patients who submitted to follow-up examinations were examined using a combined MSCT protocol of the chest and the abdomen on 4-raw MSCT (Siemens, Erlangen, Germany). For each examination, approximately 120 mL of contrast agent (Bracco-Altana, Konstanz, Germany) was applied intravenously. The raw data were transferred to an external personal computer equipped with an image reconstruction software (CardioRecon 6; Siemens, Forchheim, Germany) to simulate 5 different dose levels. To simulate them correctly, a milliampere second-dependent noise was added to every image, so that the changes in a current-time product could be imitated. The images were compared in consensus by 2 radiologists who were not informed about the technical scanning parameters, that is, dose parameters, and were graded in 4 different subcategories on a 1-to-5-point scale. For statistical analysis, the Friedman test was used. Additional evaluations for lesions smaller than 1.0 cm were performed and analyzed separately.

Results: For image noise, there was a significant change between 40 and 60 mA s. For lesion detection, there was no significant change. The contour of the small objects did not differ between 90, 60, and 40 mA s. However, a dose reduction to 30 mA s had a significant effect. The contrast did not differ between 90, 60, 40, and even down to 30 mA s. Only a maximal dose reduction to 20 mA s had a significant effect on the contrast. The level of noise was most sensitive to the current. Whereas a dose reduction to 60 mA s did not yet have a significant effect, there was a significant increase of noise at 40 mA s.

Conclusions: The MSCT can be applied with a lower dose than that usually selected in examinations at present to follow-up and stage the oncological patients adequately.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RCT.0b013e3181c95b67DOI Listing

Publication Analysis

Top Keywords

dose reduction
12
oncological patients
8
image quality
8
reduction contrast
8
dose
7
image
5
simulated low-dose
4
low-dose computed
4
computed tomography
4
tomography oncological
4

Similar Publications

Donafenib is an improved version of sorafenib in which deuterium is substituted into the drug's chemical structure, enhancing its stability and antitumor activity. Donafenib exhibits enhanced antitumor activity and better tolerance than sorafenib in preclinical and clinical studies. However, the specific mechanism of its effect on hepatocellular carcinoma has not been reported.

View Article and Find Full Text PDF

The treatment regimen for [Lu]Lu-prostate-specific membrane antigen (PSMA) 617 therapy follows that of chemotherapy: 6 administrations of a fixed activity, each separated by 6 wk. Mathematic modeling can be used to test the hypothesis that the current treatment regimen for a radiopharmaceutical modality is suboptimal. A mathematic model was developed to describe tumor growth during [Lu]Lu-PSMA therapy.

View Article and Find Full Text PDF

The tumor microenvironment characterized by heterogeneously organized vasculatures causes intra-tumoral heterogeneity of oxygen partial pressure at the cellular level, which cannot be measured by current imaging techniques. The intra-tumoral cellular heterogeneity may lead to a reduction of therapeutic effects of radiation. The purpose of this study was to investigate the effects of the heterogeneity on biological effectiveness of H-, He-, C-, O-, and Ne-ion beams for different oxygenation levels, prescribed dose levels, and cell types.

View Article and Find Full Text PDF

Unraveling the impact of PFOA toxicity on Zostera marina using a multi-omics approach: Insights from growth, physiological, transcriptomic, and metabolomic signatures.

J Hazard Mater

December 2024

Key Laboratory of Mariculture (Ocean University of China), Ministry of Education, Qingdao, PR China; Joint Research Center for Conservation, Restoration & Sustainable Utilization of Marine Ecology, Ocean University of China-China State Shipbuilding Corporation Environmental Development Co., Ltd., Qingdao, PR China; Observation and Research Station of Yellow-Bohai Sea Temperate Seagrass Bed Ecosystem, Ministry of Natural Resources, Qingdao, PR China. Electronic address:

Perfluorooctanoic acid (PFOA), an anthropogenic organic pollutant known for its persistence, resistance to degradation, and toxicity, has raised significant concerns about its potential ecological impacts. Zostera marina, a common submerged seagrass species in temperate offshore areas, is highly vulnerable to pollutant stressors. However, the impact of PFOA on Z.

View Article and Find Full Text PDF

Objectives: The purpose of this study was to investigate the fundamental properties of spot-scanning proton beams and compare them to Monte Carlo (MC) simulations, both with and without CT calibration, using spatially diverse combinations of materials.

Methods: A heterogeneous phantom was created by spatially distributing titanium, wax, and thermocol to generate six scenarios of heterogeneous combinations. Proton pencil beams ranging in energy from 100 to 226.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!