Purpose: To develop and evaluate a technique for measuring the radial resistive force, chronic outward force, and dimensions of self-expanding stents.

Materials And Methods: A Mylar film was looped around the stent, threaded through two carbon fiber rods, and immersed in a 37 degrees C oil bath. A force gauge mounted on a micro-positioning stage was used to measure the applied forces. The apparatus containing the self-expanding nitinol stent (diameter, 40 mm; length, 80 mm) was placed inside a micro-computed tomographic (CT) scanner. At each stent deformation, the load was manually recorded from the force gauge and a micro-CT volume (isotropic voxel spacing, 0.15 mm) obtained. Stent diameter and length were measured from the images, and radial resistive force and chronic outward force were calculated for each deformation.

Results: The stress-strain curves indicate that the stents exert much smaller maximum outward forces (1.2 N/cm) than the force that is required to compress them (3.6 N/cm). The forces were measured with a precision of +/-3.3% (standard deviation of five repeated measurements). The stent's diameter was measured with precision better than 0.3% and accuracy of +/-0.1 mm.

Conclusions: The authors have developed a radiographic technique that enables precise measurements of radial resistive force, chronic outward force, and the dimensions of self-expanding stents during deformation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvir.2010.01.016DOI Listing

Publication Analysis

Top Keywords

radial resistive
12
resistive force
12
force chronic
12
chronic outward
12
outward force
12
force
9
technique measuring
8
force dimensions
8
dimensions self-expanding
8
force gauge
8

Similar Publications

Per- and polyfluoroalkyl substances (PFAS) are environmentally persistent contaminants that are often referred to as "Forever Chemicals". They are used in industrial and household products; however, they are resistant to degradation. Thus, PFAS contamination has become a wide-spread issue.

View Article and Find Full Text PDF

Wearable motion capture gloves enable the precise analysis of hand and finger movements for a variety of uses, including robotic surgery, rehabilitation, and most commonly, virtual augmentation. However, many motion capture gloves restrict natural hand movement with a closed-palm design, including fabric over the palm and fingers. In order to alleviate slippage, improve comfort, reduce sizing issues, and eliminate movement restrictions, this paper presents a new low-cost data glove with an innovative open-palm and finger-free design.

View Article and Find Full Text PDF

Objective: The use of vertical margin design in all-ceramic restoration has generated inquiries regarding its clinical efficacy under diverse dynamic oral conditions. This research aims to assess the marginal fit and fracture resistance of monolithic zirconia crowns featuring vertical margin design as opposed to those with conventional horizontal margin design.

Materials And Methods: Two metal dies were employed to generate replicated resin dies mimicking mandibular first molar preparation.

View Article and Find Full Text PDF

Introduction: The aim of the present study was to investigate the effects of eccentric phase tempo in squats on hypertrophy, strength, and contractile properties of the quadriceps femoris (QF) muscle.

Methods: Eighteen participants (10 males and 8 females, age 24.0 ± 1.

View Article and Find Full Text PDF

Complex traits influenced by multiple genes pose challenges for marker-assisted selection (MAS) in breeding. Genomic selection (GS) is a promising strategy for achieving higher genetic gains in quantitative traits by stacking favorable alleles into elite cultivars. Resistance to Fusarium oxysporum f.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!