[Roseland® prosthesis: quality of life's studies about 68 patients with a mean followed-up of 43.8 months].

Chir Main

Service de chirurgie de la main du Pr-Legré, CHU de la Conception, Marseille, France.

Published: October 2010

Objectives: Roseland® prosthesis is a ball and socket prosthesis, physiological and not anatomical. This study wants to demonstrate by a quality life questionnaire (quick DASH described by Dubert et al., 2001) that Roseland® prosthesis gives to patients a trapeziometacarpal joint native capacities.

Patients And Methods: An exterior examinator reviewed prospectively 68 patients having a mean age of 61.1 years at surgery with 11 bilateral cases that is 79 prothesis. With a mean follow-up of 43.8 months, we value by Kapandji's opposition, first comisssural openning, quality of life and patient's satisfaction.

Results: Three patients have been excuded: two of them had got post-traumatic dislocation: one trapezium fracture, one unknown reason. The third patient had got osteophytis with "came" effects. We keep 65 patients with 84,6% satisfying and very satisfaying. 75,4% of patients have a capacity of 80% and more of their joint, 40% of them had got 100%.

Discussion: Roseland® prosthesis has good results because it agrees with already known principles as on its own concept: rotula prothesis gives three axes mobility, as on its own conception: metarcarpal stem with a palmar "T" shaped against rotation and bone saving, a cup with equatorial ring to prevent burying of the spongy bone. Componenents are recovered by hydroxyapatite to favour osteo-integration with less loosening than ciment. An accurate technique avoids dislocations traps: trapezium implant centring, internal osteophytis removal. An accurate indication: trapeziometarpal joint osteoarthritis only (second degree's Dell classification).

Conclusion: Roseland® prosthesis reproduces a satisfactory and functional joint for 84,6% of cases. These good results can be obtained by accurate indication (Dell II) and contra-indication (osteoarthritis around trapezium except trapeziometacarpal of course).

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.main.2010.01.005DOI Listing

Publication Analysis

Top Keywords

roseland® prosthesis
16
quality life
8
accurate indication
8
patients
6
prosthesis
5
[roseland® prosthesis
4
prosthesis quality
4
quality life's
4
life's studies
4
studies patients
4

Similar Publications

Due to its availability and biocompatibility, the human amniotic membrane (hAM) is being investigated by a large number of researchers with the goal of gaining a better understanding of the materials' mechanical behavior and structural integrity and optimizing them for various Tissue Engineering applications. In this research, biopolymers sodium alginate (SA) and silk fibroin (SF) were electrospun onto a decellularized hAM, resulting in two types of hybrid scaffolds: hAM/SF and hAM/SF/SA. The mechanical characteristics of these nanofibers were then analyzed to guide scaffold optimization for applications using these materials.

View Article and Find Full Text PDF

Biliary duct reconstruction is one of the most challenging parts of liver transplantation and accounts for 40%-60% of complications. While current stent-based devices on the market show promising results in reducing complications, they are manufactured from permanent synthetic materials and require a second reintervention for their removal. This exposes the patients to other potential complications and increases healthcare costs.

View Article and Find Full Text PDF

Does the Learning Curve for Intraoral Scanning Vary Depending on the Device?

Eur J Prosthodont Restor Dent

January 2025

Department of Dental Materials and Prosthesis, Ribeirão Preto School of Dentistry, University of São Paulo (USP), Ribeirão Preto, São Paulo, Brazil.

Objectives: Digital impressions using intraoral scanners (IOS) are replacing conventional impressions, requiring a learning process. This study compared the learning curves of 29 dentists with no prior IOS experience, randomly assigned to groups: Eagle, Omnicam- AF, and IS-3700.

Methods: After a lecture, participants performed three full scans (maxilla, mandible, and occlusion record) on a phantom and completed a survey about their experience.

View Article and Find Full Text PDF

Electrotactile stimulation is a çmethod of activating the tactile system by bypassing cutaneous mechanoreceptors and exciting the cutaneous afferent endings directly. This method is of interest for its potential in wearable tactile augmentation technologies. The generation of meaningful electrotactile sensation could benefit cases of peripheral neuropathy or prosthesis.

View Article and Find Full Text PDF

Peri-implant diseases, such as peri-implantitis, affect up to 47% of dental implant recipients, primarily due to biofilm formation. Current decontamination methods vary in efficacy, prompting interest in polymeric nanoparticles (NPs) for their antimicrobial and protein-specific cleaning properties. This study evaluated the efficacy of polymeric nanoparticles (NPs) in decontaminating titanium dental implants by removing proteinaceous pellicle layers and resisting recontamination.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!