Background: To assess the reliability of the measurements obtained with the PalmScan, when compared with another standardized A-mode ultrasound device, and assess the consistency and correlation between the two methods.

Methods: Transversal, descriptive, and comparative study. We recorded the axial length (AL), anterior chamber depth (ACD) and lens thickness (LT) obtained with two A-mode ultrasounds (PalmScan A2000 and Eye Cubed) using an immersion technique. We compared the measurements with a two-sample t-test. Agreement between the two devices was assessed with Bland-Altman plots and 95% limits of agreement.

Results: 70 eyes of 70 patients were enrolled in this study. The measurements with the Eye Cubed of AL and ACD were shorter than the measurements taken by the PalmScan. The differences were not statistically significant regarding AL (p < 0.4) but significant regarding ACD (p < 0.001). The highest agreement between the two devices was obtained during LT measurement. The PalmScan measurements were shorter, but not statistically significantly (p < 0.2).

Conclusions: The values of AL and LT, obtained with both devices are not identical, but within the limits of agreement. The agreement is not affected by the magnitude of the ocular dimensions (but only between range of 20 mm to 27 mm of AL and 3.5 mm to 5.7 mm of LT). A correction of about 0.5 D could be considered if an intraocular lens is being calculated. However due to the large variability of the results, the authors recommend discretion in using this conversion factor, and to adjust the power of the intraocular lenses based upon the personal experience of the surgeon.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2856532PMC
http://dx.doi.org/10.1186/1471-2415-10-8DOI Listing

Publication Analysis

Top Keywords

eye cubed
12
a-mode ultrasound
8
transversal descriptive
8
descriptive comparative
8
comparative study
8
measurements palmscan
8
agreement devices
8
palmscan
5
measurements
5
comparison biometric
4

Similar Publications

Purpose: To assess the accuracy and performance of the Butterfly IQ+ handheld ultrasound (HHUS) in detecting retinal detachments (RDs).

Methods: A cross-sectional observational study of eyes with (n = 20) and without (n = 20) RD imaged using a conventional B-scan ophthalmic ultrasound device Ellex Eye Cubed by Clarion Medical Technologies (Cambridge, ON, CA) and a portable HHUS (Butterfly IQ+). Images were compared between the modalities for qualitative differences.

View Article and Find Full Text PDF

Objective: Near-point of convergence (NPC) testing is an attractive screening tool in the sport setting because it is rapid, requires few resources, and is easy to administer. Remote NPC has been reported after sport-related concussion (SRC), although the incidence among a university-aged population is not well defined. The purpose of the study was to examine the incidence of remote NPC after SRC in a cohort of Canadian interuniversity athletes.

View Article and Find Full Text PDF

Purpose: To assess the rate of change in macular drusen load in fellow eyes of patients with unilateral neovascular age-related macular degeneration (AMD) to evaluate whether the change in drusen load determines the onset and type of choroidal neovascularisation (CNV).

Methods: Subjects with unilateral neovascular AMD with a minimum of 2 years follow-up were identified retrospectively from the hospital electronic database. Drusen count, area and volume measurements at the macula of the contralateral eye were recorded using the commercial software on the Topcon 3D OCT-2000 devices over the previous 2 years.

View Article and Find Full Text PDF

The Effect of Pulsing on Transverse Ultrasound Efficiency and Chatter.

Am J Ophthalmol

November 2017

Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah. Electronic address:

Purpose: To evaluate the effects of micropulse, long pulse, and continuous ultrasound on transverse ultrasound using Abbott Medical Optics' (AMO) WhiteStar Signature Pro with the Ellips FX handpiece.

Design: In vitro laboratory study.

Methods: This study was conducted at the John A.

View Article and Find Full Text PDF

EX-PRESS® Implant Position and Function: Comparative Evaluation with Ultrasound Biomicroscopy and Optical Coherence Tomography.

Middle East Afr J Ophthalmol

August 2016

Department of Ophthalmology, University Hospital of Heraklion, Heraklion, Greece; Department of Refractive Surgery, Institute of Vision and Optics, Crete, Greece.

Purpose: This study evaluated the feasibility of anterior segment optical coherence tomography (OCT) and ultrasound biomicroscopy (UBM) for the imaging of EX-PRESS(®) implant.

Materials And Methods: This nonrandomized comparative case series was performed at the Department of Ophthalmology of the University Hospital of Heraklion, Crete, Greece. The Ellex Eye Cubed (40 MHz) UBM and the Zeiss Visante OCT systems were used.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!