Objective: The examination of the urine remains to be one of the most commonly performed tests in laboratory practice. Currently, laboratories also need to accredit their urine diagnostics by comparing their measurement methods to acceptable references. In this study we compared particle counts obtained by new generation automated technique, image capture analysis (IQ-200) with those of a standardized chamber counts.

Design And Methods: The same 258 urine samples from different departments of a hospital assayed by IQ-200 were analyzed in parallel with the KOVA cell chamber system. Clinically significant discrepancy results (positive vs. negative) for red blood cell (RBC) and white blood cell (WBC) were also compared with those obtained by dipstick testing.

Results: There was a good agreement between the automated system and sediment microscopy for RBCs, WBCs, and squamous epithelial cells (SCs) (r=0.90; r=0.80; r=0.72, respectively: P<0.001). The IQ-200 was more sensitive for determining RBCs, WBCs, and SCs than other formed elements.

Conclusions: IQ-200 can perform accurate quantification of microscopic element in urine. However, automated techniques are not completely free of error. Therefore, by adopting an appropriate algorithm and combining the results with stript analysis and other laboratory tests allows further reduction of clinically important errors.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6647716PMC
http://dx.doi.org/10.1002/jcla.20319DOI Listing

Publication Analysis

Top Keywords

kova cell
8
cell chamber
8
blood cell
8
generation iq-200
4
iq-200 automated
4
urine
4
automated urine
4
urine microscopy
4
microscopy analyzer
4
analyzer compared
4

Similar Publications

Effect of the technical variability of counting chambers upon the interpretation of sperm concentration results.

Reprod Biomed Online

May 2024

External Quality Assurance Schemes for Reproductive Medicine, Northlands, Western Australia, Australia.

Research Question: Does the choice of sperm-counting chamber affect the proportion of samples generating results with an erroneous interpretation?

Design: Laboratories in an external quality assurance programme were sent 141 semen samples over a 12-year period and asked to return the sperm concentration and whether or not the result was abnormal. Only those using 5th edition of the World Health Organization manual (WHO5) interpretation criteria were included. Submissions from specialist fertility laboratories were used to calculate assigned values for each sample.

View Article and Find Full Text PDF

Background: The IDEXX SediVue Dx (SediVue) is an automated, in-clinic urine sediment analyzer for veterinary patients. The bias between the results from manual microscopy and the SediVue is currently unknown.

Objectives: To assess the diagnostic accuracy of the SediVue, we aimed to determine the bias between the SediVue (index test) and manual microscopy (reference standard) for the quantification of RBCs and WBCs in urine.

View Article and Find Full Text PDF

In this study, we examined the in vivo toxicity of the liposomes F consisting of 1,26-bis(cholest-5-en-3-yloxycarbonylamino)-7,11,16,20-tetraazahexacosan tetrahydrochloride, lipid-helper 1,2-dioleoyl-sn-glycero-3-phosphoethanolamine and folate lipoconjugate (-{2-[-2,3-di(tetradecyloxy)prop-1-yloxycarbonyl]aminoethyl}-'-[2-(pteroyl-L-glutam-5-yl)aminoethyl]octadecaethyleneglycol) and investigated the antitumor effect of combined antitumor therapy consisting of MDR1-targeted siMDR/F complexes and conventional polychemotherapy using tumor xenograft initiated in immunodeficient mice. Detailed analysis of acute and chronic toxicity of this liposomal formulation in healthy C57BL/6J mice demonstrated that formulation F and parent formulation L (without folate lipoconjugate) have no acute and chronic toxicity in mice. The study of the biodistribution of siMDR/F lipoplexes in SCID mice with xenograft tumors formed by tumor cells differing in the expression level of folate receptors showed that the accumulation in various types of tumors strongly depends on the abandons of folate receptors in tumor cells and effective accumulation occurs only in tumors formed by cells with the highest FR levels.

View Article and Find Full Text PDF

Performance of four chambers to measure sperm concentration: results from an external quality assurance programme.

Reprod Biomed Online

October 2020

External Quality Assurance Schemes for Reproductive Medicine, PO Box 162, Northlands Western Australia 6905, Australia.

Research Question: What are the changes in the use of four types of counting chambers by laboratories enrolled in an Australian-based external quality assurance programme, and what are their accuracy and precision?

Design: Samples of latex beads of known concentration up to 20 × 10/ml were distributed quarterly to enrolled laboratories over a 12-year period. The results of each distribution were then used to calculate a bias relative to the target value as an indicator of accuracy and a coefficient of variation to indicate the level of precision.

Results: The proportion of laboratories in 2007-2008 using improved Neubauer haemocytometers (44%), Makler® (9%) and Vetriplast chambers (19%) remained constant in 2018-2019, unlike Kova chamber users (20%), which reduced.

View Article and Find Full Text PDF

[Performance analysis of the Sysmex UF4000/UD10 for diagnosis of urinary tract infections].

Ann Biol Clin (Paris)

December 2019

Assistance publique hôpitaux de Paris, Laboratoire de microbiologie clinique, Hôpital Necker-Enfants Malades, Paris, France.

Background: we evaluated the performance of the flow cytometry-based UF-4000 automated urine analyzer associated with the UD-10 image analyzer (Sysmex), in comparison with optical microscopy and culture.

Materials And Methods: 2,695 consecutive urine samples of patients were included. The cell count was performed using the analyzer and the Kova cell for 316 samples, and compared according to a threshold of 10 white blood cells (WBC) and 10 red blood cells (RBC) /μL.

View Article and Find Full Text PDF