Background: Predischarge defibrillation threshold testing is often performed a few days after ICD implantation in order to validate defibrillation thresholds obtained at the time of implant. Ventricular fibrillation is induced with such testing and causes an increase in serum Brain Natriuretic Peptide (BNP) levels. BNP is an indicator for cardiac stress. We wanted to examine the feasibility to alter the trend of BNP after predischarge testing in VVI, DDD and CRT ICD's.
Methods: We measured BNP before predischarge testing and 5, 10, 20 and 40 minutes after predischarge testing in 13 groups with each 20 patients. We evaluated patients without post shock pacing and patients with a post shock pacing frequency of 60, 70, 80, 90 and 100 bpm and a duration of 30 and 60 sec as well as a post shock pacing frequency of 80 and 90 bpm and a duration of 120 sec post shock pacing.
Results: Patients without post shock pacing showed the highest BNP during the follow-up. The percentage values of BNP increased consistent significantly after 5 minutes compared with BNP before predischarge testing. The percentage values of BNP trend was significantly lower with a post shock pacing of 90 bpm and duration of 60 sec. In addition, we excluded a cardiac necrosis by predischarge testing because of similar values of myoglobin, cardiac troponin I and creatine kinase during the follow-up.
Conclusions: Our results suggested that post shock pacing with 90 bpm and duration of 60 sec as the best optimized post shock pacing frequency and duration for VVI, DDD and CRT ICD's. A reduction of cardiac stress is going to be achieved with the optimization of the post shock pacing frequency and duration.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2836008 | PMC |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!