Contact angle measurements are used to infer the clinical wetting characteristics of contact lenses. Such characterization has become more commonplace since the introduction of silicone hydrogel contact lens materials, which have been associated with reduced in vivo wetting due to the inclusion of siloxane-containing components. Using consistent methodology and a single investigator, advancing and receding contact angles were measured for 11 commercially available silicone hydrogel contact lens types with a dynamic captive bubble technique employing customized, fully automated image analysis. Advancing contact angles were found to range between 20° and 72° with the lenses falling into six statistically discrete groupings. Receding contact angles fell within a narrower range, between 17° and 22°, with the lenses segregated into three groups. The relationship between these laboratory measurements and the clinical performance of the lenses requires further investigation.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0885328210363505DOI Listing

Publication Analysis

Top Keywords

silicone hydrogel
12
hydrogel contact
12
contact angles
12
contact angle
8
contact
8
contact lenses
8
contact lens
8
receding contact
8
lenses
5
dynamic contact
4

Similar Publications

Inspired by counterintuitive water "swelling" ability of the hydrophobic moss of the genus Sphagnum (Peat moss), we prepared a hydrophobic pseudo-hydrogel (HPH), composed of a pure hydrophobic silicone elastomer with a tailored porous structure. In contrast to conventional hydrogels, HPH achieves absorption-induced volume expansion through surface tension induced elastocapillarity, presenting an unexpected absorption-induced volume expansion capability in hydrophobic matrices. We adopt a theoretical framework elucidating the interplay of surface tension induced elastocapillarity, providing insights into the absorption-induced volume expansion behavior.

View Article and Find Full Text PDF

The discharge of oil-laden wastewater from industrial processes and the frequent occurrence of oil spills pose severe threats to the ecological environment and human health. Membrane materials with special wettability have garnered attention for their ability to achieve efficient oil-water separation by leveraging the differences in wettability at the oil-water interface. These materials are characterized by their simplicity, energy efficiency, environmental friendliness, and reusability.

View Article and Find Full Text PDF

Low-cost male urogenital simulator for penile implant surgery training: a 3D printing approach.

3D Print Med

January 2025

Department of Surgical & Interventional Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.

Background: Penile implant surgery is the standard surgical treatment for end-stage erectile dysfunction. However, the growing complexity of modern high-tech penile prostheses has increased the demand for more practical training opportunities. The most advanced contemporary training methods involve simulation training using cadavers, with costs exceeding $5,000 per cadaver, inclusive of biohazard fees.

View Article and Find Full Text PDF

Quarter of a century of contact lens prescribing trends in Australia (2000-2024).

Clin Exp Optom

January 2025

Eurolens Research, Division of Pharmacy and Optometry, The University of Manchester, Manchester, UK.

Clinical Relevance: Knowledge of contact lens prescribing trends can (a) assist practitioners to benchmark their own prescribing habits, (b) help the contact lens industry understand preferred products, and (c) support academics in scholarly writings.

Background: This work aims to document contact lens prescribing trends in Australia over the past quarter of a century.

Methods: An annual survey of contact lens prescribing trends was conducted in Australia each year from 2000 to 2024, inclusive, by asking optometrists to provide information relating to 10 consecutive contact lens fits undertaken between January and March.

View Article and Find Full Text PDF

Objectives: To compare the efficacy of two different silicone hydrogel bandage contact lenses (BCLs) in terms of visual rehabilitation and ocular discomfort following photorefractive keratectomy (PRK).

Materials And Methods: This prospective study included 60 eyes of 30 patients who underwent bilateral PRK surgery to correct myopia and/or astigmatism refractive errors. Following surgery, lotrafilcon A BCLs were applied to the right eye and senofilcon A BCLs were applied to the left eye.

View Article and Find Full Text PDF