Background: Radio Frequency Identification (RFID) devices are becoming more and more essential for patient safety in hospitals. The purpose of this study was to determine patient safety, data reliability and signal loss wearing on skin RFID devices during magnetic resonance imaging (MRI) and computed tomography (CT) scanning.

Methods: Sixty RFID tags of the type I-Code SLI, 13.56 MHz, ISO 18000-3.1 were tested: Thirty type 1, an RFID tag with a 76 x 45 mm aluminum-etched antenna and 30 type 2, a tag with a 31 x 14 mm copper-etched antenna. The signal loss, material movement and heat tests were performed in a 1.5 T and a 3 T MR system. For data integrity, the tags were tested additionally during CT scanning. Standardized function tests were performed with all transponders before and after all imaging studies.

Results: There was no memory loss or data alteration in the RFID tags after MRI and CT scanning. Concerning heating (a maximum of 3.6 degrees C) and device movement (below 1 N/kg) no relevant influence was found. Concerning signal loss (artifacts 2 - 4 mm), interpretability of MR images was impaired when superficial structures such as skin, subcutaneous tissues or tendons were assessed.

Conclusions: Patients wearing RFID wristbands are safe in 1.5 T and 3 T MR scanners using normal operation mode for RF-field. The findings are specific to the RFID tags that underwent testing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2825188PMC
http://dx.doi.org/10.1186/1754-9493-4-2DOI Listing

Publication Analysis

Top Keywords

signal loss
12
rfid tags
12
radio frequency
8
frequency identification
8
devices magnetic
8
magnetic resonance
8
resonance imaging
8
computed tomography
8
rfid devices
8
patient safety
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!