Unlabelled: We have developed a methodology for calibrating (68)Ge radioactivity content in a commercially available calibration source for activity calibrators in a way that is traceable to the national standard. Additionally, the source was cross-calibrated for equivalent (18)F content by direct comparison with the national standard for (18)F in the same geometry.

Methods: Sources containing standardized (68)GeCl(4) or (18)F-FDG solutions were prepared at the National Institute of Standards and Technology (NIST) with mock syringe blanks used in the construction of a commercially available epoxy-based (68)Ge calibration source. These sources and several NIST-constructed epoxy-based (68)Ge mock syringes were then used as artifact standards to determine calibration factors for NIST-maintained activity calibrators and secondary standard ionization chambers to enable calibration of the actual commercial sources. A direct comparison between the solution-based (68)Ge sources and the (18)F-FDG sources allowed for an empiric determination of the relative response for these radionuclides in several commercial activity calibrators. Potential measurement effects due to differences between the solution composition and the epoxy and theoretic (68)Ge-to-(18)F response ratios were studied by Monte Carlo simulation.

Results: The calibration factors developed in this study enabled NIST to calibrate epoxy-based mock syringe sources with a relative combined standard uncertainty of 0.52%. The direct comparisons of the (68)Ge and (18)F standards in the various ionization chambers allowed the activity to be expressed in terms of equivalent (18)F activity with a relative combined standard uncertainty of about 0.9%.

Conclusion: The ability for NIST to calibrate these epoxy-based mock syringes enabled, for the first time to our knowledge, the direct traceability to the national (68)Ge standard to be established for this type of source. Through a direct comparison with the NIST (18)F standard, the determination of the relative response ratios in activity calibrators enabled the equivalent (18)F activity to be determined in a way that was also traceable to the national (18)F activity standard.

Download full-text PDF

Source
http://dx.doi.org/10.2967/jnumed.109.070300DOI Listing

Publication Analysis

Top Keywords

activity calibrators
20
equivalent 18f
12
direct comparison
12
18f activity
12
activity
9
18f
8
calibration source
8
traceable national
8
standard
8
national standard
8

Similar Publications

BuShao Tiaozhi Capsule (BSTZC), a novel drug in China, has been used to treat hyperlipidemia (HLP) in clinical practice for many years. Despite our previous studies suggesting that BSTZC can treat HLP, there is a lack of a rapid and systematic method to explore its active components. Therefore, in this study, we aimed to investigate the active components and mechanisms of BSTZC in treating HLP by integrating serum pharmacology, pharmacokinetics, network analysis, and experimental validation.

View Article and Find Full Text PDF

Unlabelled: Calcium imaging is a key method to record the spiking activity of identified and genetically targeted neurons. However, the observed calcium signals are only an indirect readout of the underlying electrophysiological events (single spikes or bursts of spikes) and require dedicated algorithms to recover the spike rate. These algorithms for spike inference can be optimized using ground truth data from combined electrical and optical recordings, but it is not clear how such optimized algorithms perform on cell types and brain regions for which ground truth does not exist.

View Article and Find Full Text PDF

Comprehensive Qualitative Drug Screening in Emergency Toxicology Using an Automated LC-MS System:Simultaneous Quantification of Relevant Drugs and Metabolites in Blood Plasma.

Drug Test Anal

January 2025

Department of Experimental and Clinical Toxicology, Institute of Experimental and Clinical Pharmacology and Toxicology, Center for Molecular Signaling (PZMS), Saarland University, Homburg, Germany.

Rapid and comprehensive qualitative and quantitative analytical procedures are crucial in 24/7 emergency toxicology (ET) to support diagnosis and treatment of acute intoxications and to monitor their progression and efficacy of detoxification strategies. This study aimed to develop the simultaneous qualitative and quantitative analysis of 62 drugs, as well as seven active metabolites in human blood plasma using an automated liquid chromatography (LC)-linear ion trap mass spectrometry (MS) screening system. Sample preparation was conducted by liquid-liquid extraction, and plasma concentrations were determined using an electronically stored 5-point calibration.

View Article and Find Full Text PDF

Purpose: The LIMOS is a validated observational tool designed to assess stroke rehabilitation within hospital settings. To enable its application in outpatient settings in Brazil and similar low- and middle-income countries, it is necessary to adapt LIMOS into a Patient-Reported Outcome Measure (PROM). Accordingly, this study aimed to translate, cross-culturally adapt, and standardise LIMOS for the Brazilian context (LIMOS-Br) and transform it into a PROM.

View Article and Find Full Text PDF

Biological age estimation from DNA methylation and determination of relevant biomarkers is an active research problem which has predominantly been tackled with black-box penalized regression. Machine learning is used to select a small subset of features from hundreds of thousands of CpG probes and to increase generalizability typically lacking with ordinary least-squares regression. Here, we show that such feature selection lacks biological interpretability and relevance in the clocks of the first and next generations and clarify the logic by which these clocks systematically exclude biomarkers of aging and age-related disease.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!