Object: In this study, long-term results are presented from clinical studies of the Bryan Cervical Disc Prosthesis at University Hospital Gasthuisberg in Leuven, Belgium. A total of 98 patients (89 with 1-level and 9 with 2-level implantations) agreed to participate in follow-up studies for up to 10 years postoperatively. This article focuses on the 4- and 6-year results. Patients in one of the clinical studies had either radiculopathy or myelopathy associated with spondylosis and/or disc herniations that did not respond to conservative treatment. Patients from the other clinical study received commercially available Bryan devices and the study protocol did not have specific inclusion/exclusion criteria. More than 90% of the patients were considered to have radiculopathy.

Methods: Clinical measurements discussed in the article include the 36-Item Short Form Health Survey, Neck Disability Index, numerical ratings of neck and arm pain, neurological outcomes, and Odom classification. Angular motion findings from lateral flexion-extension radiographs are also presented. The occurrence of adverse events and second surgeries are examined as an indicator of device safety.

Results: The clinical outcomes at 4 and 6 years postoperatively appear consistent with the previously reported results at 1 and 2 years postoperatively. The mean angular motion results at 4 and 6 years postoperatively for 1-level patients were 7.3 and 7.7 degrees, respectively. Two-level patients had slightly less motion at 4 and 6 years postoperatively with mean caudad values of 5.7 and 6.0 degrees, respectively, and cephalad values of 4.2 and 6.2 degrees, respectively. Efforts were made to capture adverse events, regardless of their nature and relatedness to the study surgery. This effort resulted in a relatively high number of recorded events. However, only 6 patients experienced events that were judged by the investigator to be related, either possibly or definitely, to the Bryan device. These events included device migration, device removal, and hoarseness and vocal cord paralysis, as well as 3 cases involving pain and neurological symptoms. Eight patients underwent further neck surgery to treat symptoms.

Conclusions: The favorable clinical and angular motion outcomes that were previously noted at 1- and 2-years' follow-up after cervical disc replacement with the Bryan Cervical Disc Prosthesis appear to persist after 4 and 6 years of follow-up.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2009.9.SPINE09129DOI Listing

Publication Analysis

Top Keywords

cervical disc
20
years postoperatively
20
bryan cervical
12
disc prosthesis
12
angular motion
12
follow-up cervical
8
disc replacement
8
clinical studies
8
patients
8
patients clinical
8

Similar Publications

Comparison of Recruitment Method on Clinical Outcomes Following Cervical Disc Arthroplasty.

Spine (Phila Pa 1976)

January 2025

Indiana Spine Group Location of investigation Indiana Spine Group, 13225 N. Meridian Street, Carmel, IN 46032.

Study Design: Retrospective cohort.

Objective: To compare the clinical outcomes of trial versus standard clinical practice (SCP) patients following cervical disc arthroplasty (CDA).

Background: CDA is hypothesized to reduce the shear strain and related complications resulting from fusion procedures.

View Article and Find Full Text PDF

Transoral resection of a symptomatic odontoid process aneurysmal bone cyst: illustrative case.

J Neurosurg Case Lessons

January 2025

Departments of Neurosurgery, NYU Langone Health, New York, New York.

Background: Aneurysmal bone cysts (ABCs) are slow-growing, expansile bone tumors most often observed in the long bones and lumbar and thoracic spine. Anterior column ABCs of the spine are rare, and few cases have described their surgical management, particularly for lesions with extension into the odontoid process and the bilateral C2 pedicles. In the present case, the authors describe a two-stage strategy for resection of a symptomatic 2.

View Article and Find Full Text PDF

Background: Intervertebral disc (IVD) degeneration is the main cause of neck pain. Although conventional magnetic resonance imaging can detect morphological changes in intervertebral disc degeneration, it cannot provide accurate and objective evaluations. Magnetic resonance diffusion tensor imaging (DTI) reflects the microstructural changes in tissues by describing the diffusion of water molecules.

View Article and Find Full Text PDF

Study Design: Retrospective study.

Objective: We aim to determine the influence of preoperative Patient-Reported Outcome Measurement Information System Physical Function (PROMIS-PF) scores on perioperative and postoperative outcomes, the latter determined through patient-reported outcome measures (PROMs) and the degree of achievement rates of minimum clinically important difference (MCID) following single-level cervical disc replacement (CDR).

Background: Several studies have examined the relationship between preoperative PROMIS-PF as a prognostic factor for postoperative outcomes.

View Article and Find Full Text PDF

Clinical characteristics associated with cervical hydrated nucleus pulposus extrusion in dogs.

J Vet Intern Med

January 2025

Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, New York, USA.

Background: Clinical characteristics of cervical hydrated nucleus pulposus extrusion (HNPE) in dogs compared to other causes of cervical myelopathy are not well described.

Hypothesis/objectives: To evaluate for clinical characteristics and mechanical ventilation likelihood associated with HNPE compared to other causes of cervical myelopathy.

Animals: Three hundred seventy-seven client-owned dogs from 2010 to 2022.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!