Objective: To compare the biomechanics of costotransverse process screw fixation with those of pedicle screw fixation in a cadaveric model of the upper thoracic spine.

Methods: Ten human thoracic spines were instrumented across the T3-T4 segment with costotransverse and pedicle screws. Nonconstraining pure moments (maximum, 6.0 Nm) were applied to induce flexion, extension, lateral bending, and axial rotation. The range of motion, lax zone, and stiff zone were determined in each specimen in the normal state, after 3-column destabilization, and after instrumentation. After flexibility testing was completed, axial screw pull-out strength was assessed.

Results: In all directions of loading, both fixation techniques significantly decreased lax zone and range of motion at T3-T4 compared with the destabilized state (P < .001). During all loading modes except lateral bending, pedicle screw fixation allowed significantly less range of motion than costotransverse screw fixation. Pedicle screws provided 62% greater resistance to axial pull-out than costotransverse screws.

Conclusion: The costotransverse screw technique seems to provide only moderately stiff fixation of the destabilized thoracic spine. Pedicle screw fixation seems to have more favorable biomechanical properties. These data suggest that the costotransverse process construct is better used as a salvage procedure rather than as a primary fixation strategy.

Download full-text PDF

Source
http://dx.doi.org/10.1227/01.NEU.0000350869.35779.05DOI Listing

Publication Analysis

Top Keywords

screw fixation
28
pedicle screw
16
costotransverse process
12
fixation pedicle
12
range motion
12
fixation
10
screw
9
process screw
8
upper thoracic
8
thoracic spine
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!