We illustrate how the use of heated electrodes enhances the performance of glucose biosensors based on amperometric detection of the glucose-oxidase generated hydrogen peroxide. Nafion is shown to be an excellent matrix to protect glucose oxidase from thermal inactivation during the heating pulses. The influence of the electrode temperature upon the amperometric response is examined. Temperature pulse amperometry (TPA) has been used to obtain convenient peak-shaped analytical signals. Surprisingly, up to 67.5 °C, the activity of Nafion-entrapped glucose oxidase is greatly enhanced (24 -fold) by accelerated kinetics rather than decreased by thermal inactivation. Amperometric signals even at elevated temperatures are stable upon prolonged operation involving repetitive measurements. The linear calibration range is significantly extended.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2745167 | PMC |
http://dx.doi.org/10.1016/j.elecom.2009.07.030 | DOI Listing |
Biosensors (Basel)
January 2025
Graduate Program in Chemistry, Federal University of Maranhão, São Luís 65080-805, MA, Brazil.
The technology of 3D printing, particularly fused deposition modeling (FDM) 3D printing, has revolutionized the development of electrochemical biosensors, offering a versatile and cost-effective approach for clinical applications. This review explores the integration of FDM in fabricating biosensing platforms tailored for clinical diagnostics, emphasizing its role in detecting various biomarkers and viral pathogens. Advances in 3D printing materials, especially the emergence of bespoke conductive filaments, have allowed the production of highly customizable and efficient biosensors.
View Article and Find Full Text PDFBiosensors (Basel)
January 2025
Faculty of Civil Engineering and Mechanics, Kunming University of Science and Technology, Kunming 650500, China.
Continuous glucose monitoring based on the minimally invasive implantation of glucose sensor is characterized by high accuracy and good stability. At present, glucose concentration monitoring based on fluorescent glucose capsule sensor is a new development trend. In this paper, we design a fluorescent glucose capsule sensor with a design optimization study.
View Article and Find Full Text PDFBiosensors (Basel)
December 2024
Cofoe Medical Technology Co., Ltd., No. 816 Zhenghua Road, Changsha 410021, China.
Electrochemical glucose sensor holds significant promise for the monitoring of blood glucose levels in diabetic patients. In this study, we proposed a novel electrochemical glucose sensor based on 1,10-Phenanthroline-5,6-dione (PD)/Ru(III) as a dual redox mediator. The synergistic effect of PD and Ru(III) was utilized to efficiently facilitate the electron transfer between the enzyme-active center and the electrode.
View Article and Find Full Text PDFLancet Diabetes Endocrinol
January 2025
Department of Cardiovascular Disease, Saint Luke's Mid America Heart Institute, University of Missouri-Kansas City School of Medicine, Kansas City, MO, USA. Electronic address:
Background: About half of patients with heart failure with mildly reduced or preserved ejection fraction (HFpEF) have type 2 diabetes. In the STEP-HFpEF DM trial of adults with obesity-related HFpEF and type 2 diabetes, subcutaneous once weekly semaglutide 2·4 mg conferred improvements in heart failure-related symptoms and physical limitations, bodyweight, and other heart failure outcomes. We aimed to determine whether these effects of semaglutide differ according to baseline HbA.
View Article and Find Full Text PDFJ Mater Chem B
January 2025
Departamento de Química Inorgánica, Orgánica y Bioquímica, Facultad de Ciencias y Tecnologías Químicas, Universidad de Castilla-La Mancha, 13071 Ciudad Real, Spain.
Diabetes remains one of the most prevalent chronic diseases globally, significantly impacting mortality ratetables. The development of effective treatments for controlling glucose level in blood is critical to improve the quality of life of patients with diabetes. In this sense, smart optical sensors using hydrogels, responsive to external stimuli, have emerged as a revolutionary approach to diabetes care.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!