Safety of oxidized zirconium (OxZr) in total knee arthroplasty (TKA) has been supported by biomechanical, clinical, and radiologic data. Retrieved OxZr femoral components and corresponding polyethylene (PE) inserts were examined to rule out patterns leading to early failure. Sixteen retrieved TKA with an OxZr femoral component were included. The PE inlays were analyzed applying an established scoring system for wear and surface damage. Femoral components were screened for scratching, pitting, delamination, and striations. The prostheses were in situ for 16.4 ± 11.9 months. The average wear and damage score for the tibial PE inserts was 36.0 ± 12.7. The average score in the visual analysis of the OxZr femoral components was 1.3 ± 1.3. The data presented in this study did not show major wear of the PE in TKA using OxZr components. In this cohort, there were no failures directly related to this implant.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.arth.2009.11.024 | DOI Listing |
Acta Orthop
September 2024
Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway.
Background And Purpose: We aimed to evaluate polyethylene (PE) wear, cup migration, and clinical outcome over 10 years in total hip arthroplasties (THA) using different articulations.
Methods: This is a secondary analysis of 150 patients randomized into 5 groups, using different articulations: Charnley/Charnley Ogee for steel and conventional polyethylene (CPE), or Spectron EF/Reflection with either CPE or highly cross linked polyethylene (XLPE) cups, paired with heads made of either cobalt-chromium (CoCr) or oxidized zirconium (OxZr). All cups were cemented.
Knee
December 2023
Lower Limb Arthroplasty Unit, Liverpool University Hospitals NHS Foundation Trust, L14 3LB, United Kingdom; Faculty of Health and Life Sciences, The University of Liverpool, L69 3BX, United Kingdom. Electronic address:
Background: Early studies have found Oxidised Zirconium (OxZr-Nb) total knee arthroplasties to have a low incidence of failure in young, high demand patients. Theoretically this is because they are low friction and hard wearing. However, there are a paucity of studies reporting on the outcome of these implants beyond ten years.
View Article and Find Full Text PDFJ Mech Behav Biomed Mater
August 2023
Laboratory of Biomechanics and Implant Research, Department of Orthopaedics, Heidelberg University Hospital, Schlierbacher Landstrasse 200a, 69118, Heidelberg, Germany.
Background: In total knee arthroplasty, femoral components with coated or oxidized surfaces are commonly used as alternative to CoCrMo in metal sensitive patients. Data on the in vivo behaviour of different coating types is, however, rare. Aim of the study was the investigation of coating stability with respect to implant and patient specific parameters.
View Article and Find Full Text PDFBone Joint J
July 2022
Department of Trauma and Orthopaedics, University College Hospital, London, UK.
Aims: This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA).
Methods: Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups.
J Arthroplasty
August 2022
Department of Orthopedic Surgery, Washington University School of Medicine, St. Louis, Missouri.
Background: A study was performed to measure metal ions present in the knee joint after performing a total knee arthroplasty (TKA) with standard cobalt chromium (CoCr) components as well as with "nickel-free" oxidized zirconium femoral and titanium tibial (OxZr/Ti) components.
Methods: Knee joint fluid was collected prior to arthrotomy, and on postoperative day one to determine the amount of metal debris generated when performing a TKA with standard instrumentation from consecutive cases with CoCr components (n = 24) and OxZr/Ti components (n = 16).
Results: CoCr implant patients had statistically higher levels of nickel (Ni) (29.