Pilot evaluation of a prototype critical care blood glucose monitor in normal volunteers.

J Diabetes Sci Technol

Division of Research, Department of Anesthesiology, Cooper University Hospital, The Robert Wood Johnson Medical School-UMDNJ, Camden, New Jersey 08103, USA.

Published: November 2009

Background: Availability of a highly accurate in-hospital automated blood glucose (BG) monitor could facilitate implementation of intensive insulin therapy protocols through effective titration of insulin therapy, improved BG control, and avoidance of hypoglycemia. We evaluated a functional prototype BG monitor designed to perform frequent automated blood sampling for glucose monitoring.

Methods: Sixteen healthy adult volunteer subjects had intravenous catheter insertions in a forearm or hand vein and were studied for 8 hours. The prototype monitor consisted of an autosampling unit with a precise computer-controlled reversible syringe pump and a glucose analytical section. BG was referenced against a Yellow Springs Instrument (YSI) laboratory analyzer. Sampling errors for automated blood draws were assessed by calculating the percent of failed draws, and BG data were analyzed using the Bland and Altman technique.

Results: Out of 498 total sample draws, unsuccessful draws were categorized as follow: 11 (2.2%) were due to autosampler technical problems, 21 (4.2%) were due to catheter-related failures, and 37 (7.4%) were BG meter errors confirmed by a glucometer-generated error code. Blood draw difficulties or failures related to the catheter site (e.g., catheter occlusion or vein collapse) occurred in 6/15 (40%) subjects. Mean BG bias versus YSI was 0.20 +/- 12.6 mg/dl, and mean absolute relative difference was 10.4%.

Conclusions: Automated phlebotomy can be performed in healthy subjects using this prototype BG monitor. The BG measurement technology had suboptimal accuracy based on a YSI reference. A more accurate BG point-of-care testing meter and strip technology have been incorporated into the future version of this monitor. Development of such a monitor could alleviate the burden of frequent BG testing and reduce the risk of hypoglycemia in patients on insulin therapy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2787022PMC
http://dx.doi.org/10.1177/193229680900300602DOI Listing

Publication Analysis

Top Keywords

automated blood
12
insulin therapy
12
prototype monitor
12
blood glucose
8
glucose monitor
8
monitor
7
blood
5
pilot evaluation
4
prototype
4
evaluation prototype
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!