Background: We sought to test the feasibility and efficacy of bihormonal closed-loop blood glucose (BG) control that utilizes subcutaneous (SC) infusion of insulin and glucagon, a model-predictive control algorithm for determining insulin dosing, and a proportional-derivative control algorithm for determining glucagon dosing.

Methods: Thirteen closed-loop experiments (approximately 7-27 h in length) were conducted in six ambulatory diabetic pigs weighing 26-50 kg. In all experiments, venous BG was sampled through a central line in the vena cava. Efficacy was evaluated in terms of the controller's ability to regulate BG in response to large meal disturbances ( approximately 5 g of carbohydrate per kilogram of body mass per meal) based only on regular frequent venous BG sampling and requiring only the subject's weight for initialization.

Results: Closed-loop results demonstrated successful BG regulation to normoglycemic range, with average insulin-to-carbohydrate ratios between approximately 1:20 and 1:40 U/g. The total insulin bolus doses averaged approximately 6 U for a meal containing approximately 6 g per kilogram body mass. Mean BG values in two 24 h experiments were approximately 142 and approximately 155 mg/dl, with the total daily dose (TDD) of insulin being approximately 0.8-1.0 U per kilogram of body mass and the TDD of glucagon being approximately 0.02-0.05 mg. Results also affirmed the efficacy of SC doses of glucagon in staving off episodic hypoglycemia.

Conclusions: We demonstrate the feasibility of bihormonal closed-loop BG regulation using a control system that employs SC infusion of insulin and glucagon as governed by an algorithm that reacts only to BG without any feed-forward information regarding carbohydrate consumption or physical activity. As such, this study can reasonably be regarded as the first practical implementation of an artificial endocrine pancreas that has a hormonally derived counterregulatory capability.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2769971PMC
http://dx.doi.org/10.1177/193229680900300428DOI Listing

Publication Analysis

Top Keywords

bihormonal closed-loop
12
infusion insulin
12
insulin glucagon
12
kilogram body
12
body mass
12
closed-loop blood
8
blood glucose
8
glucose control
8
subcutaneous infusion
8
ambulatory diabetic
8

Similar Publications

Objectives: Impaired awareness of hyperglycaemia (IAH) affects approximately 20-40% of people living with type 1 diabetes (T1D), predisposing them to severe hypoglycaemia. This systematic review evaluated the efficacy of closed-loop automated insulin delivery systems (CL-AID) in restoring IAH compared with standard diabetes care, including other diabetes technologies.

Methods: Six electronic databases were searched for published and unpublished observational and randomised-control studies (RCTs) from inception to 29th of May 2024.

View Article and Find Full Text PDF

Background: Management of insulin administration for intake of carbohydrates and physical activity can be burdensome for people with type 1 diabetes on hybrid closed-loop systems. Bihormonal fully closed-loop (FCL) systems could help reduce this burden. In this trial, we assessed the long-term performance and safety of a bihormonal FCL system.

View Article and Find Full Text PDF

Introduction: The management of type 1 diabetes (T1DM) has undergone significant advancements with the availability of novel technologies, notably continuous and flash glucose monitoring (CGM and FGM, respectively) and hybrid closed loop (HCL) therapy. The dual hormone fully closed loop (DHFCL) approach with insulin and glucagon infusion has shown promising effects in small studies on glycaemic regulation and quality of life in T1DM.

Methods And Analysis: The Dual Hormone Fully Closed Loop for Type 1 Diabetes (DARE) study is a non-commercial 12-month open-label, two-arm randomised parallel-group trial.

View Article and Find Full Text PDF

Introduction: This review aims to critically appraise differences in methodology and quality of model-based and empirical-data-based cost-utility studies to address key limitations, opportunities, and challenges to inform future cost-utility analyses of continuous glucose monitoring (CGM) in type 1 diabetes. This protocol is registered at PROSPERO (CRD42023391284).

Methods: The review will be conducted in accordance with the PRISMA guideline for systematic reviews.

View Article and Find Full Text PDF

Currently, continuous glucose monitoring sensors are used in the artificial pancreas to monitor blood glucose levels. However, insulin and glucagon concentrations in different parts of the body cannot be measured in real-time, and determining body glucagon sensitivity is not feasible. Estimating these states provides more information about the current system status, facilitating improved decision-making by the model-based controller.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!