The non-modulated transfer (NMT) of the total effluent of the separation of polychlorinated biphenyls (PCBs) on two columns coupled in series has been studied. A non-polar poly(5%-phenyl-95%-methyl)siloxane column (40 m x 100 microm x 0.1 microm) and a polar 70%-cyanopropyl-polysilphenylene-siloxane column (4 m x 0.1 mm x 0.1 microm) were used as (1)D and (2)D columns, respectively. The effluents from both the (1)D column and the (1)D+(2)D column series were monitored independently by two FIDs. Data from the (1)D and (1)D+(2)D integration reports were used to evaluate the NMT experiment. (1)D retention times, t(R,i,D1), were directly accessible from (1)D integration report while (2)D retention times, t(R,i,B), were calculated for all corresponding peak pairs from (1)D and (1)D+(2)D integration reports as a difference t(R,i,D2)=t(R,i,D1+D2)-t(R,i,D1). Search for corresponding peaks of PCB congeners in the (1)D and (1)D+(2)D chromatograms is elucidated in detail on standard PCB samples and on PCB congeners present in the technical formulation Arochlor 1242. Both retention times (t(R,i,D1) and t(R,i,D2)) as well as peak widths at half height (w(h,i)) and peak heights (h(i)) obtained from integration reports were used to construct 2D and 3D images for PCB NMT separations on serially coupled columns. The performance of the NMT procedure is illustrated by the separation of (i) standard PCB solutions, (ii) a mixture of the 209 PCBs, and (iii) a mixture of Arochlor 1242 and hydrocarbons on the DB-5+BPX-70 column series.