Background: Measurement of anterior segment parameters plays an important role in diagnosis and treatment of glaucoma. The objective of this study was to evaluate the repeatability and reproducibility of anterior chamber volume (ACV) measurements with rotating scheimpflug camera (RSC) and to examine agreement with anterior segment optical coherence tomography (AS-OCT).

Methods: Thirty nine healthy normal subjects were recruited from the Eye Center of Tongren Hospital. ACV was measured using RSC and AS-OCT in a randomly selected eye for each subject. For RSC measurements, both automatic and manual ACV measurements and 2 independent operators' ACV measurements were obtained. All subjects were invited for 3 visits within a week to evaluate repeatability and reproducibility of ACV measurement by RSC. Agreement was evaluated between RSC and AS-OCT.

Results: Good repeatability and reproducibility were found for both automatic and manual ACV measurements obtained by RSC. For intrasession repeatability, coefficient of variation (CVw) and intraclass correlation coefficient (ICC) values for automatic were 3.52% and 0.98; the values for manual were 3.44% and 0.97, respectively. For intersession reproducibility, the respective CVw and ICC values were 3.96% and 0.96. Good agreement was also found in 2 operators for both automatic and manual ACV measurements; nevertheless, poor agreement was found between RSC and AS-OCT (95% confidence interval (CI) for agreement of automatic RSC measurement versus AS-OCT were -96.3 to 72.8 microl and 95% CI for agreement of manual RSC measurement versus AS-OCT were between -41.7 to 10.1 microl).

Conclusions: Both RSC automatic and manual ACV measurements showed good repeatability and reproducibility, and showed comparable agreement between 2 independent operators, but poor agreement was found between RSC and AS-OCT.

Download full-text PDF

Source

Publication Analysis

Top Keywords

acv measurements
24
repeatability reproducibility
16
automatic manual
16
manual acv
16
anterior segment
12
rsc as-oct
12
rsc
11
measurement anterior
8
anterior chamber
8
chamber volume
8

Similar Publications

Background: Muscarinic receptor agonism and positive allosteric modulation is a promising mechanism of action for treating psychosis, not present in most D2R-blocking antipsychotics. Xanomeline, an M1/M4-preferring agonist, has shown efficacy in late-stage clinical trials, with more compounds being investigated. Therefore, we aim to synthesize evidence on the preclinical efficacy of muscarinic receptor agonists and positive allosteric modulators in animal models of psychosis to provide unique insights and evidence-based information to guide drug development.

View Article and Find Full Text PDF

Objective: To describe demographics, causative pathogens, hospitalization, mortality, and antimicrobial resistance of bacterial bloodstream infections (BSIs) among beneficiaries in the global U.S. Military Health System (MHS), a single-provider healthcare system with 10-year longitudinal follow-up.

View Article and Find Full Text PDF

Purpose: This study evaluates the discrepancies between ACV measurements obtained from the Heidelberg Anterion and Zeiss IOLMaster 700 and investigates the significance of ACV and other ocular biometry parameters.

Patients And Methods: To investigate intraocular fluid circulation, a robust formula was developed for ACV measurement using the Zeiss IOLMaster 700. A pilot study was conducted to validate this formula, which relied on WTW, CCT, and ACD.

View Article and Find Full Text PDF

Clinical predictor of volume hypoperfusion in patients treated with extended window mechanical thrombectomy: an external validation study.

Medicina (B Aires)

December 2024

Departamento de Neurorradiología Intervencionista, Instituto Médico ENERI, Clínica La Sagrada Familia, Buenos Aires, Argentina.

Introduction: Treatment of ischemic stroke has expanded beyond the 6-hour window, but identification of eligible patients requires advanced imaging, which is often unavailable. This study aimed to validate a prediction model using the NIHSS score (a measure of stroke severity) to estimate hypoperfusion volume and determine eligibility for DEFUSE 3 treatment criteria.

Materials And Methods: Data from ischemic stroke patients with less than 24 hours' evolution were analyzed.

View Article and Find Full Text PDF

The aim of this study was to determine the validity, reliability, and sensitivity of a new linear position transducer (LPT) device (RepOne) to a previously validated LPT (Tendo) during the barbell back squat and bench press exercises. Fourteen recreationally resistancetrained individuals (7 males and 7 females) performed three repetitions for the back squat and bench press at loads ranging from 30-90% 1RM. Both devices recorded average (ACV) and peak (PCV) concentric velocities concurrently for every repetition at each load.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!