Flexibility and trackability of laser cut coronary stent systems.

Acta Bioeng Biomech

Budapest University of Technology and Economics, Department of Materials Science and Engineering, Budapest, Hungary.

Published: October 2010

Coronary stents are the most important supports in present day cardiology. Flexibility and trackability are two basic features of stents. In this paper, four different balloon-expandable coronary stent systems were investigated mechanically in order to compare their suitability. The coronary stent systems were assessed by measurements of stent flexibility as well as by comparison of forces during simulated stenting in a self-investigated coronary vessel model. The stents were cut by laser from a single tube of 316L stainless steel or L-605 (CoCr) cobalt chromium alloy. The one-and four-point bending tests were carried out to evaluate the stent flexibility E x I (Nmm(2)), under displacement control in crimped and expanded configurations. The flexibility of stents would be rather dependent on the design than on raw material. In general a more flexible stent needs lower tracking force during the implantation. The L-605 raw material stents need lower track force to pass through in the vessel model than the 316L raw material stents. The sort and long stents passed through the curved vessel model in different ways. The long stents nestled to the vessel wall at the outer arc and bent, while the short stents did not bend in the curve, only the delivery systems bent.

Download full-text PDF

Source

Publication Analysis

Top Keywords

coronary stent
12
stent systems
12
vessel model
12
raw material
12
stents
9
flexibility trackability
8
stent flexibility
8
material stents
8
long stents
8
stent
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!