Background: The use of radiofrequency identification (RFID) systems is expanding and highlights the need to address electromagnetic interference (EMI) to implantable pacemakers and implantable cardioverter-defibrillators (ICDs).

Objective: This study sought to examine the electromagnetic compatibility (EMC) between RFID readers and implantable pacemakers or ICDs.

Methods: During in vitro testing, 15 implantable pacemakers and 15 ICDs were exposed to 13 passive RFID readers in 3 frequency bands: 134 kHz (low frequency [LF]), 13.56 MHz (high frequency [HF]), and 915 MHz (ultra high frequency [UHF]).

Results: While being exposed to LF RFID, a reaction was observed for 67% of all pacemaker tests (maximum distance 60 cm) and 47% of all ICD tests (maximum distance 40 cm). During HF RFID exposure, a reaction was observed for 6% of all pacemaker tests (maximum distance 22.5 cm) and 1% of all ICD tests (maximum distance 7.5 cm). For both pacemakers and ICDs, no reactions were observed during exposure to UHF RFID or continuous-wave RFID. Pacemakers and ICDs were most susceptible to modulated LF RFID readers.

Conclusion: Although there is in vitro testing evidence for concern for implantable pacemaker and ICD EMI at LF and HF, the FDA has not received any incident reports of pacemaker or ICD EMI caused by any RFID system. We do not believe the current situation reveals an urgent public health risk.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2009.09.071DOI Listing

Publication Analysis

Top Keywords

implantable pacemakers
16
tests maximum
16
maximum distance
16
pacemakers icds
12
rfid
9
radiofrequency identification
8
electromagnetic interference
8
pacemakers implantable
8
implantable cardioverter-defibrillators
8
rfid readers
8

Similar Publications

Risk factors for long-term severe tricuspid regurgitation following mitral valve replacement: a retrospective study.

BMC Cardiovasc Disord

January 2025

Department of Cardiology, Xuzhou Central Hospital, No.199 Jiefang South Road, Quanshan District, Xuzhou, 221009, People's Republic of China.

Background: The aim of this study is to identify factors associated with the development of long-term severe tricuspid regurgitation (TR) following mitral valve replacement (MVR).

Methods: A retrospective analysis was conducted involving 308 patients who underwent single-valve MVR at Xuzhou Central Hospital between April 2017 and December 2022. Preoperative color Doppler ultrasound indicated that all patients had either no or mild to moderate tricuspid regurgitation.

View Article and Find Full Text PDF

Objective: We present our experience with endovascular Bentall procedure (Endo-Bentall) using a modular valve conduit (Endo-Bentall) in high-risk patients with aortic root pathologies.

Methods: The physician constructed Endo-Bentall device is composed of a self-expanding transcatheter aortic valve (TAVR), aortic endovascular stent graft (TEVAR), and two wire-reinforced fenestrations for coronary artery stenting. The TAVR valve is sutured into an appropriately sized TEVAR graft.

View Article and Find Full Text PDF

Background: Permanent Pacemaker (PPM) implantation is essential in treating cardiac arrhythmias and conduction disorders, especially in patients with heart failure. Although PPM has been proven to improve quality of life and prolong life expectancy in patients with cardiac conduction disorders, post implantation complications still often occur.

Purpose: This study aimed to identify types of complications and associated predictors in patients undergoing PPM implantation.

View Article and Find Full Text PDF

Cardiac strangulation with chronic ab-extrinseco occlusion of the left-circumflex artery from an epicardial lead: a case report.

Eur Heart J Case Rep

January 2025

Department of Radiological and Hematological Sciences, Section of Radiology, Università Cattolica del Sacro Cuore, Largo Agostino Gemelli 8 - 00168 Rome, Italy.

Background: Cardiac strangulation (CS) from epicardial pacing leads (EPLs) is a rare and potentially lethal mechanical complication associated with epicardial pacemaker (PM) implantation.

Case Summary: We report a case of a 44-year-old-female patient presenting with chest and left shoulder pain in the absence of reported trauma with history of congenital atrioventricular block treated with epicardial PM implantation during the childhood and subsequent transvenous reimplantation over the years. Troponin I resulted within normal values and ECG, transthoracic echocardiography and chest X-ray documented no acute cardiopulmonary findings.

View Article and Find Full Text PDF

Background: Removal of cardiac implantable electronic devices (CIEDs) is strongly recommended for CIED-related infections, and leadless pacemakers (LPs) are increasingly used for reimplantation. However, the optimal timing and safety of LP implantation after CIED removal for infection remains unclear.This systematic review and meta-analysis aimed to assess complication rates (all-cause mortality and reinfection) when LP implantation was performed simultaneously with or after CIED removal.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!