A customized modular reference array clamp for navigated spine surgery.

Arch Orthop Trauma Surg

Department of Orthopaedic Surgery, Hospital for Special Surgery, Computer Assisted Surgery Center, 535 East 70th Street, New York, NY 10021, USA.

Published: December 2010

AI Article Synopsis

  • Developed a modular fixation system for precise pedicle screw placement in spinal surgeries.
  • Conducted a cadaver study implanting 16 pedicle screws across cervical, thoracic, and lumbar regions, verifying placement accuracy with advanced imaging techniques.
  • Results showed successful screw placements without technical issues or significant image artifacts, highlighting the effectiveness of the reference array clamp system in spinal navigation.

Article Abstract

Introduction: The current authors have developed a modular system of reference array fixation which is tailored specifically to the spinal level being operated upon. They believe that this system may further increase the precision and accuracy of pedicle screw placement.

Materials And Methods: Two formalin-fixed whole body cadavers were used for this study. For cervical spine evaluation of the reference clamp, four odontoid screws (two per cadaver) for C1/C2-fusion and four lateral mass screws (two per cadaver) were implanted. Following navigated screw placement with 2D and 3D fluoroscopic verification, insertion of two lateral mass screws was performed. In the same way, lumbar and thoracic pedicle screws were implanted. Two pedicle screws were placed at two levels of the lumbar and two levels of the thoracic areas giving an overall of 16 screws implanted (8 cervical, 4 thoracic, and 4 lumbar). Postoperative evaluation involved comparison of postoperative 3D scans and preoperative planning images. A simple classification system was used for evaluation of any deviation from the planned trajectory.

Results: All pedicle screw placements were performed as planned without any technical problems. The reference array clamps remained in position at all the spinal levels at which they were employed with no loosening or displacement and no secondary damage to any of the spinous processes. Manual manipulation was performed but no displacement or slippage was observed. Image artefacts caused by the reference clamp were not significant as to obscure the area of interest. Both imaging modalities (Iso-C 3D and Vario 3D) generated sufficiently precise 3D images. There was no substantial difference in quality when those two systems were compared.

Discussion: Insufficient fixation of the reference clamp can lead to failure and complications. To date, no reference clamp systems have been developed specifically for navigated spine surgery.

Conclusions: Stable reference array fixation is a critical step in navigated surgery. To date, the same reference clamps have been applied to the spinal anatomy as have been developed originally for the appendicular skeleton. The current investigators have developed a novel modular clamp and have demonstrated its efficacy in a cadaveric model.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00402-010-1060-0DOI Listing

Publication Analysis

Top Keywords

reference array
16
reference clamp
16
reference
9
navigated spine
8
array fixation
8
pedicle screw
8
screws cadaver
8
lateral mass
8
mass screws
8
pedicle screws
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!