Accuracy and precision of manufacturer-supplied liquid medication administration devices before and after patient education.

J Am Pharm Assoc (2003)

Bernard J. Dunn School of Pharmacy, Shenandoah University, 1775 North Sector Court, Winchester, VA 22601, USA.

Published: May 2010

Objectives: To compare the accuracy and precision of a manufacturer-supplied dropper versus a pharmacy-supplied syringe when measured by a trained student pharmacist and to evaluate patients' ability to accurately measure a prescribed amount of medication with these devices before and after counseling on their use.

Methods: This was a two-phased study with a pretest/posttest design of 125 community pharmacy patrons of Virginia community pharmacies. Deliverable volume was measured by a student pharmacist for nystatin suspension and digoxin elixir using the manufactured-supplied device. Patients at pharmacies were asked to complete a short survey, then asked to measure a specified dose of nystatin suspension and digoxin elixir (using the manufacturer-supplied administration device). The patients repeated this a second time after a student provided education on how to properly use each administration device. The main outcomes measure was whether education on an administration device would provide more measurement accuracy using medications with different physical properties.

Results: For nystatin, 88% and 85.6% of patients measured with an error of greater than or equal to 20% before and after counseling, respectively. For digoxin, the totals were 24.8% before counseling and 4% after. A statistically significant (P < 0.05) difference was observed in the percent measurement error before and after counseling for nystatin and digoxin both for patients who had been educated before the study and for those who had not been counseled.

Conclusion: In general, the accuracy of medication dosage devices seems to be most affected by viscosity of the liquid being measured. With education on proper use, the oral syringe provides a statistically and clinically significant improvement in dosing accuracy for viscous medications compared with the manufacturer-supplied dropper. Pharmacists must review with patients the proper way to use a manufacturer- or pharmacy-supplied oral delivery device for appropriate dosing measurements.

Download full-text PDF

Source
http://dx.doi.org/10.1331/JAPhA.2010.09006DOI Listing

Publication Analysis

Top Keywords

administration device
12
accuracy precision
8
precision manufacturer-supplied
8
manufacturer-supplied dropper
8
student pharmacist
8
nystatin suspension
8
suspension digoxin
8
digoxin elixir
8
device patients
8
accuracy
5

Similar Publications

Sacubitril/valsartan (Sac/Val) belongs to the group of angiotensin receptor-neprilysin inhibitors and has been used for the treatment of heart failure (HF) for several years. The mechanisms that mediate the beneficial effects of Sac/Val are not yet fully understood. In this study we investigated whether Sac/Val influences the two proteolytic systems, the ubiquitin-proteasome system (UPS) and the autophagy-lysosomal pathway (ALP), in a mouse model of pressure overload induced by transverse aortic constriction (TAC) and in human induced pluripotent stem cell-derived cardiomyocytes (hiPSC-CMs) treated with endothelin-1 (ET1) serving as a human cellular model of hypertrophy.

View Article and Find Full Text PDF

Introduction: The internet of things (IoT) is increasingly used for occupational safety and health (OSH) purposes in private and public organisations. Current practices and regulations are unclear, and some stakeholders raised concerns about deploying this technology at work.

Methods: Following the PRISMA-ScR checklist, we reviewed the main opportunities and ethical issues raised by using IoT devices for OSH purposes, as discussed in the academic literature.

View Article and Find Full Text PDF

The study was aimed at assessing the role of the MiniMed780G system of glycemic control before, during, and after Ramadan among people with Type 1 diabetes (PwT1D). This is a single-center retrospective analysis of MiniMed780G system users aged 14 years and above whose glycemic profiles were collected from February 21 to May 20, 2023, which corresponds to the Hijri months of Sha'ban, Ramadan, and Shawwal 1444/1445. Data was collected, processed, and analyzed in the framework of the Medtronic Galaxy service of the One Hospital Clinical Service (OHCS) program in Dallah Hospital, Riyadh, Saudi Arabia.

View Article and Find Full Text PDF

Omecamtiv mecarbil in precision-cut living heart failure slices: A story of a double-edged sword.

J Mol Cell Cardiol Plus

September 2023

Translational Cardiothoracic Surgery Research Lab, Department of Cardiothoracic Surgery, Erasmus University Medical Center, Rotterdam, the Netherlands.

Heart failure (HF) is a rapidly growing pandemic while medical treatment options remain limited. Omecamtiv mecarbil (OM) is a novel HF drug that directly targets the myosin heads of the cardiac muscle. This study used living myocardial slices (LMS) from patients with HF to evaluate the direct biomechanical effects of OM as compared to dobutamine.

View Article and Find Full Text PDF

The delivery of molecules, such as DNA, RNA, peptides, and certain hydrophilic drugs, across the epidermal barrier poses a significant obstacle. Microneedle technology has emerged as a prominent area of focus in biomedical research because of its ability to deliver a wide range of biomolecules, vaccines, medicines, and other substances through the skin. Microneedles (MNs) form microchannels by disrupting the skin's structure, which compromises its barrier function, and facilitating the easy penetration of drugs into the skin.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!