The purpose of this study was to assess the effect of a driven equilibrium (DRIVE) pulse incorporated in a standard T1-weighted turbo spin echo (TSE) sequence as used in our routine MRI protocol for examination of pathologies of the knee. Sixteen consecutive patients with knee disorders were examined using the routine MRI protocol, including T1-weighted TSE-sequences with and without a DRIVE pulse. Signal-to-noise ratios (SNRs) and contrast-to-noise ratio (CNR) of anatomical structures and pathologies were calculated and compared for both sequences. The differences in diagnostic value of the T1-weighted images with and without DRIVE pulse were assessed. SNR was significantly higher on images acquired with DRIVE pulse for fluid, effusion, cartilage and bone. Differences in the SNR of meniscus and muscle between the two sequences were not statistically significant. CNR was significantly increased between muscle and effusion, fluid and cartilage, fluid and meniscus, cartilage and meniscus, bone and cartilage on images acquired using the DRIVE pulse. Diagnostic value of the T1-weighted images was found to be improved for delineation of anatomic structures and for diagnosing a variety of pathologies when a DRIVE pulse is incorporated in the sequence. Incorporation of a DRIVE pulse into a standard T1-weighted TSE-sequence leads to significant increase of SNR and CNR of both, anatomical structures and pathologies, and consequently to an increase in diagnostic value within the same acquisition time.