Objective: To perform a systematic review of validation studies of blood pressure measuring devices done using the European Society of Hypertension International Protocol (ESH-IP) since its publication in 2002.
Methods: Major aspects of ESH-IP studies are described. A review of the ESH-IP performance, problems and violations in its application, and the effect of applying several more stringent validation criteria in an ESH-IP revision is carried out.
Results: From January 2002 to June 2009, 104 validation studies had been conducted using the ESH-IP, 36 using the British Hypertension Society protocol and 28 using the US Association for the Advancement of Medical Instrumentation protocol. Among 78 studies reported up to June 2008, in 66 (85%) the tested device has passed the ESH-IP. In 19 validations a modification of the ESH-IP was performed to adapt for specific study needs (population or device). Protocol violations of the ESH-IP were identified in 23 studies (eight major violations). A test of several arbitrarily chosen changes in the ESH-IP validation criteria applied in the published studies showed the phase 2.1 criterion (BP differences < or =15 mmHg) and the phase 2.2 criteria to be the more stringent.
Conclusion: The ESH-IP has succeeded in expanding the validation procedure worldwide by three to four-fold compared with the period before its publication. There is a need for protocol revision aiming to address issues that appeared in published studies, prevent protocol violations, and ensure complete data reporting. Standardization of the ESH-IP validation studies' report and application of more stringent criteria should be considered.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MBP.0b013e3283360eaf | DOI Listing |
Blood Press Monit
August 2024
Nuffield Department of Primary Care Health Sciences, University of Oxford.
Objective: To report a validation of the Riester Big Ben Square Desk Aneroid Sphygmomanometer according to the international protocol developed by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension 2002 (ESH-IP 2002) in the interest of transparency. This legacy publication is intended to assure users that the device satisfied the requirements in place at that time.
Methods: Performance of the device was assessed by participants' age, sex, arm circumference and entry SBP/DBP.
PLoS One
March 2023
Department of Cardiovascular Sciences, University of Leicester, Leicester, United Kingdom.
Objective: In a large-scale population cardiovascular screening programme, peripheral artery disease (PAD) and hypertension would ideally be rapidly assessed using a single device. The ankle-brachial pressure index (ABPI) is calculated by comparing the ankle and brachial blood pressure (BP). However, it is currently unclear whether brachial BP measurements provided by automated PAD screening systems are sufficiently accurate for simultaneous hypertension screening.
View Article and Find Full Text PDFJ Clin Hypertens (Greenwich)
May 2022
Medical Outpatient Department and Hypertension Clinic, ESH Hypertension Centre of Excellence, University Hospital Basel, Basel, Switzerland.
Blood pressure monitoring (BPM) devices have to be validated according to strict international validation protocols. Each protocol requests a specific number of participants to be included. All protocols use vast amounts of resources, as three people have to be present for every measurement, making trials costly, especially when the manufacturer has no intention to execute a validation study, reflected in the low share of validated in the commercially available BPM devices.
View Article and Find Full Text PDFBlood Press Monit
December 2021
Health Management/Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University.
Background: Accurate measurement of blood pressure (BP) is crucial to hypertension control and prevention of future stroke and heart attack. All BP measuring devices must be validated independently in the clinical setting.
Objective: To validate the accuracy of three automatic upper arm devices (Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B) for self-measurement of BP in Chinese adults with arm size of 22-32 cm.
Objective: The aim of the present study was to validate the blood pressure (BP) monitor Beurer BM 28 according to the International Protocol of the European Society of Hypertension (ESH-IP) revision 2010.
Methods: In 33 subjects of age 27-81 years, BP measurements were performed according to the ESH-IP protocol, which alternates reference mercury sphygmomanometer and device-under-test (Beurer BM 28) measurements, resulting in a total of 99 comparisons.
Results: As to part 1 of the protocol, an absolute difference within 5 mmHg between the Beurer BM 28 and the test device was found in 83 out of 99 comparisons for the SBP and 82 out of 99 comparisons for the DBP.