In this paper the effect of treatment room and maze layout on the photoneutron and capture gamma dose equivalent in the maze was studied. MCNPX Monte Carlo (MC) code was used to simulate the Varian 2100 C/D Clinac 18 MV and four different room layouts. Two analytical methods, Wu-McGinley and McGinley, were used for dose calculations. The analytical methods overestimated the photoneutron dose (13-43 %) and gamma capture dose (16-95 %) comparing with the MC method at the maze entrance door. The results of MC method revealed that additional bend can cause a great reduction in photoneutron (5000 times) and capture gamma dose (50 times) in the maze entrance door.