An immunoaffinity liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed for the quantitation of the zinc endopeptidase matrix metalloproteinase 9 (MMP-9) from mouse serum. Sample preparation for the assay included magnetic bead-based enrichment using an MMP-9 antibody and was performed in a 96-well plate format using a liquid-handling robotic platform. The surrogate peptide GSPLQGPFLTAR derived from MMP-9 by trypsin digestion was monitored using an on-line capillary flow trap-release chromatography setup incorporating a series of trap columns (C18, strong cation exchange, and another C18) prior to nanoflow chromatography and nanospray ionization with selected reaction monitoring (SRM) detection. The assay was fit-for-purpose validated and found to be accurate (<15% interbatch relative error) and precise (<15% interbatch coefficient of variation) across a range from 0.03 to 7.3nM mouse MMP-9. Finally, the method was employed to measure MMP-9 concentrations in 30 naïve mouse serum samples, and results were compared with those obtained by an immunoassay.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ab.2010.01.002DOI Listing

Publication Analysis

Top Keywords

immunoaffinity liquid
8
liquid chromatography-tandem
8
chromatography-tandem mass
8
mass spectrometry
8
matrix metalloproteinase
8
mouse serum
8
spectrometry assay
4
assay quantitation
4
quantitation matrix
4
metalloproteinase mouse
4

Similar Publications

Brazil is an influential and successful food-producing country, where we can highlight artisanal cheeses gaining visibility in foreign markets. Some of these cheeses are made from raw milk, making them susceptible to contamination by microorganisms, including fungi, which can produce harmful mycotoxins. Feed contaminated with aflatoxin B1, when consumed by dairy animals, is metabolized and transformed into aflatoxin M1 (AFM1), which is excreted in milk.

View Article and Find Full Text PDF

Antibody-drug conjugates (ADCs) are intricate compounds that pose significant challenges in bioanalytical characterization. Therefore, multiple bioanalytical methods are required to comprehensively elucidate their pharmacokinetic (PK) profiles. In this study, we investigated DS001, an ADC consisting of a humanized monoclonal antibody (hRS7), a cleavable chemical linker, and the microtubule inhibitor monomethyl auristatin E (MMAE), with a drug-to-antibody ratio (DAR) of 8.

View Article and Find Full Text PDF

Objectives: An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.

View Article and Find Full Text PDF

Mycotoxins are secondary metabolites of fungi and represent a serious problem for human health. Due to growing interest, various aspects have been widely studied by scientific groups. One of these aspects relates to the food industry and associated beer production.

View Article and Find Full Text PDF

Food poisoning outbreaks frequently involve staphylococcal enterotoxins (SEs). SEs include 33 distinct types and multiple sequence variants per SE type. Various mass spectrometry methods have been reported for the detection of SEs using a conventional bottom-up approach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!