Article Abstract

Until now, the PFA-100 system has been considered unsuitable for monitoring clopidogrel efficacy. The authors evaluated platelet function in peripheral arterial occlusive disease (PAOD) patients using a new PFA-100(R) test cartridge (product name: INNOVANCE PFA P2Y*) specifically designed for this purpose. Twenty-two stable PAOD patients on antithrombotic therapy with clopidogrel alone (n = 22) and 18 patients undergoing a peripheral catheter intervention, preliminarily treated with 100 mg/day of aspirin followed by co-administration of clopidogrel (loading dose 300 mg, maintenance dose 75 mg/day), were enrolled in this study. Defining non-responsiveness to clopidogrel as an aggregation response within the reference range (90% central interval), four (18.2%) non-responders using light transmittance aggregometry (LTA) induced by 5 microM adenosine diphosphate (ADP) and six (27.3%) non-responders using LTA induced by 2 microM ADP (LateAggr >72.1% and >42.9%, respectively) were identified. INNOVANCE PFA P2Y* determined six (27.3%) non-responders (CT < 87 s). Agreement between the two aggregometry assays and INNOVANCE PFA P2Y* on the definition of clopidogrel response and non-response exceeded 70%. Only three patients were uniformly identified as clopidogrel non-responders by all three assays. When clopidogrel was co-administered with aspirin, two (11.1%) non-responders to clopidogrel were detected with INNOVANCE PFA P2Y*, whereas ADP-induced LTA found all patients to be responsive. INNOVANCE PFA P2Y* appears to be suitable for monitoring the effect of clopidogrel on platelet function. Its sensitivity in detecting responsiveness or non-responsiveness to clopidogrel is comparable to ADP-induced LTA. Additional prospective studies are needed to clarify the clinical relevance of the test results and classification obtained with INNOVANCE PFA P2Y*.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00277-009-0881-9DOI Listing

Publication Analysis

Top Keywords

innovance pfa
28
pfa p2y*
24
clopidogrel
10
pfa-100 system
8
test cartridge
8
monitoring clopidogrel
8
platelet function
8
paod patients
8
non-responsiveness clopidogrel
8
lta induced
8

Similar Publications

von Willebrand factor antigen, von Willebrand factor propeptide and ADAMTS13 activity in TIA or ischaemic stroke patients changing antiplatelet therapy.

J Neurol Sci

August 2024

Vascular Neurology Research Foundation, c/o Department of Neurology, Tallaght University Hospital (TUH) / The Adelaide and Meath Hospital, Dublin, incorporating the National Children's Hospital (AMNCH), Dublin, Ireland; Department of Neurology, TUH / AMNCH, Dublin, Ireland; Stroke Service, TUH / AMNCH, Dublin, Ireland; Department of Clinical and Movement Neurosciences, Royal Free Campus, UCL Queen Square Institute of Neurology, London, UK; Academic Unit of Neurology, School of Medicine, Trinity College Dublin, Ireland. Electronic address:

Data are limited on the impact of commencing antiplatelet therapy on von Willebrand Factor Antigen (VWF:Ag) or von Willebrand Factor propeptide (VWFpp) levels and ADAMTS13 activity, and their relationship with platelet reactivity following TIA/ischaemic stroke. In this pilot, observational study, VWF:Ag and VWFpp levels and ADAMTS13 activity were quantified in 48 patients ≤4 weeks of TIA/ischaemic stroke (baseline), and 14 days (14d) and 90 days (90d) after commencing aspirin, clopidogrel or aspirin+dipyridamole. Platelet reactivity was assessed at moderately-high shear stress (PFA-100® Collagen-Epinephrine / Collagen-ADP / INNOVANCE PFA P2Y assays), and low shear stress (VerifyNow® Aspirin / P2Y12, and Multiplate® Aspirin / ADP assays).

View Article and Find Full Text PDF

Aspirin Resistance and Ischemic Stroke.

Pril (Makedon Akad Nauk Umet Odd Med Nauki)

June 2024

2Faculty of Medical Sciences, Goce Delcev University, Stip, R.N. Macedonia.

: Acetylsalicylic acid (ASA) is the most widely used antiplatelet agent in treating ischemic strokes. ASA resistance varies between the populations, from 5% to 60%. We aimed to determine the prevalence of ASA resistance in our population, its association with demographic characteristics, risk factors, and the occurrence of recurrent ischemic stroke (IS).

View Article and Find Full Text PDF

Various abnormalities of coagulation such as primary hemostasis, secondary hemostasis and fibrinolysis have been reported in patients with subclinical and overt hypothyroidism. Platelets are major elements of primary hemostasis and endothelial repair. Platelet size, shape and number are the determinant of platelet function.

View Article and Find Full Text PDF

Devices such as stents and flow diverters require the use of safe and fast antiplatelet therapy. We aimed to compare the responses to clopidogrel, prasugrel, and ticagrelor by assessing the Platelet Function Analysis (PFA-100)-Innovance test results of patients undergoing endovascular stenting to determine their resistance rates. Sixty-one women and 55 men, aged 18-87 years, were included in this study.

View Article and Find Full Text PDF

Towards 50 years of platelet function analyser (PFA) testing.

Clin Chem Lab Med

April 2023

Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy.

The platelet function analyser (PFA) is a prevalent platelet function screening instrument, and comes in two models-the original PFA-100 and the contemporary PFA-200. The instruments have 'identical' output, being a 'closure time' (CT). Moreover, normal reference ranges provided by the manufacturer, for the specific test cartridges, are the same for both models.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!