Background And Purpose: Some recent studies on radiation lens injuries have indicated much lower dose thresholds than specified by the current radiation protection guidelines. The purpose of this research was to measure the lens dose during brain CT scans with multidetector row CT and to assess methods for estimating the lens dose.

Materials And Methods: With 8 types of multidetector row CT scanners, both axial and helical scans were obtained for the head part of a human-shaped phantom by using normal clinical settings with the orbitomeatal line as the baseline. We measured the doses on both eyelids by using an RPLGD during whole-brain scans including the orbit with the starting point at the level of the inferior orbital rim. To assess the effect of the starting points on the lens doses, we measured the lens doses by using 2 other starting points for scanning (the orbitomeatal line and the superior orbital rim).

Results: The CTDIvols and the lens doses during whole-brain CT including the orbit were 50.9-113.3 mGy and 42.6-103.5 mGy, respectively. The ratios of lens dose to CTDIvol were 80.6%-103.4%. The lens doses decreased as the starting points were set more superiorly. The lens doses during scans from the superior orbital rim were 11.8%-20.9% of the doses during the scans from the inferior orbital rim.

Conclusions: CTDIvol can be used to estimate the lens dose during whole-brain CT when the orbit is included in the scanning range.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7964184PMC
http://dx.doi.org/10.3174/ajnr.A1946DOI Listing

Publication Analysis

Top Keywords

lens doses
20
lens dose
16
lens
12
multidetector row
12
starting points
12
brain scans
8
scans multidetector
8
row scanners
8
including orbit
8
inferior orbital
8

Similar Publications

Objectives: To evaluate the effects of short-acting cycloplegic agents, tropicamide and compound tropicamide, on ocular biological parameters and choroid thickness.

Methods: In this study, seventy pediatric subjects aged 6 to 13 years were randomly assigned to two groups: the tropicamide group and compound tropicamide group. Ocular biological parameters and choroidal thickness (CT) and subfoveal choroid thickness (SFCT) were measured in both groups and were retested 40 min after drug administration.

View Article and Find Full Text PDF

A systematic review of the effectiveness of leaded glasses for ensuring safety among healthcare professionals in fluoroscopy.

J Med Imaging Radiat Sci

January 2025

Instituto Politécnico de Coimbra, ESTESC - Coimbra Health School, Medical Imaging and Radiotherapy, Rua 5 de Outubro, S. Martinho do Bispo, Coimbra 3046-854, Portugal. Electronic address:

Background: Currently, there is an increase in procedures across various clinical specialties involving the use of ionising radiation.

Objective: The primary objective of this systematic review is to analyse and compare the existing literature regarding the effectiveness of leaded glasses for healthcare professionals.

Methods: Comprehensive literature searches were conducted for relevant studies published between 2018 and 2023 using the Scopus, PubMed, and Web of Science databases according to preferred reporting items for systematic reviews and meta-analyses (PRISMA) methodology.

View Article and Find Full Text PDF

Eye lens dosimetry: does the direction of rotation (vertical or horizontal) play a role in type testing?

J Radiol Prot

January 2025

Radiation Protection Dosimetry (6.3), Physikalisch-Technische Bundesanstalt, Braunschweig, NDS, GERMANY.

With the International Commission on Radiological Protection (ICRP) lowering the annual dose limit for the eye lens to 20 mSv, precise monitoring of eye lens exposure has become essential. The personal dose equivalent at a depth of 3 mm, Hp(3), is the measurement method for monitoring the dose to the lens of the eye. Traditional dosimetry methods primarily address lateral radiation exposure scenarios, where radiation approaches from the left or right, necessitating the rotation of the phantom during type testing around the vertical axis.

View Article and Find Full Text PDF

Background: Radiographic skeletal survey plays an important role in the diagnosis of infant abuse. Some practitioners have expressed concerns about the radiation exposure from this examination.

Objective: To utilize state-of-the-art hybrid computational phantoms to more accurately estimate radiation doses of skeletal surveys performed for suspected infant abuse.

View Article and Find Full Text PDF

Characterization of the Interaction of Human γS Crystallin with Metal Ions and Its Effect on Protein Aggregation.

Biomolecules

December 2024

LABRMN, Centro de Investigaciones Químicas, Instituto de Investigación en Ciencias Básicas y Aplicadas, Universidad Autónoma del Estado de Morelos, Cuernavaca 62209, Mexico.

Cataracts are diseases characterized by the opacity of the ocular lens and the subsequent deterioration of vision. Metal ions are one of the factors that have been reported to induce crystallin aggregation. For HγS crystallin, several equivalent ratios of Cu(II) promote protein aggregation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!