Objectives: To investigate the performance of semi-automated measurements (RECIST, volume) of hepatic metastases in multidetector-row computed tomography (MDCT) under normal-dose- and simulated low-dose-protocols.

Materials And Methods: Thirty-five patients (67 +/- 13 years) with a total of 79 hepatic metastases underwent 16-MDCT (120 kv, 160 mAseff, pitch 1, 3 mm slice thickness, 2 mm reconstruction increment, B30f standard soft tissue kernel) for either initial staging or therapy monitoring. Corresponding raw data from these standard-dose scans were simulated at lower radiation doses of 80/60/40 mAseff (Somatom Noise Vers.6.1 beta, Siemens Healthcare, Forchheim, Germany). A semi-automated software tool (SyngoCT Oncology, Siemens Healthcare, Forchheim, Germany) was applied to each dose setting to evaluate size parameters (RECIST, volume). These measurements were compared by applying repeated-measures analysis of variance and displayed graphically.

Results: For RECIST measurements no statistically significant differences were found between standard dose (Mean RECIST diameter: 20.46 +/- 8.37 mm) and different simulated low radiation doses (80 mAseff: 20.95 +/- 8.20 mm/60 mAseff: 20.50 +/- 8.35 mm/40 mAseff: 19.95 +/- 8.16 mm): P = 0.0774.Statistically significant differences of volume quantification (P < 0.05) could be found between standard-(3.60 +/- 4.63 mL) and simulated lowest dose of 40 mAseff (3.17 +/- 4.08 mL), whereas there was no difference (P > 0.05) between 160 mAseff- and either 80 mAseff-(3.46 +/- 4.31 mL) or 60 mAseff-protocols (3.44 +/- 4.35 mL).

Conclusions: Software-assisted assessment of RECIST criteria and volume demonstrated valid performances under different dose-settings in MDCT; therefore, substantial radiation dose reduction could be possible with the use of semi-automated measurements in follow-up studies.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0b013e3181cd194eDOI Listing

Publication Analysis

Top Keywords

+/-
9
dose reduction
8
reduction semi-automated
8
semi-automated measurements
8
recist volume
8
hepatic metastases
8
radiation doses
8
siemens healthcare
8
healthcare forchheim
8
forchheim germany
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!