We surveyed the actual conditions of mammography with regard to image quality and radiation dose at 44 facilities in Kagoshima prefecture in 1999. In April 2004, guidelines for mammography newly included the standard of digital mammography. From September to October 2005, the survey was conducted at 48 facilities, and the results of the survey were compared with that in 1999. We visited 44 of the 48 facilities, and visually evaluated the image quality of mammograms for RMI156 and clinical mammograms. In addition, we measured average mammary gland dose at each facility. The number of the mammography device that satisfied the specified guideline criterion was larger than that in 1999. Image quality for the RMI156 mammograms improved. However, the results of the present survey revealed several problems. First, the number of facilities that had quality control instruments for mammography are few. Second, radiological technologists, medical doctors, and nurses did not share knowledge or information regarding mammography. Finally, there were differences in devices and image quality for mammography among the facilities. We achieved an understanding of the actual conditions of mammography in Kagoshima prefecture by visiting many facilities, evaluating image quality, and communicating with many staff members. Our results may be useful for the development of mammography examinations.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.65.1526DOI Listing

Publication Analysis

Top Keywords

image quality
20
actual conditions
12
conditions mammography
12
kagoshima prefecture
12
mammography
10
mammography kagoshima
8
quality
6
facilities
6
image
5
investigation actual
4

Similar Publications

Purpose: Our study evaluated skeletal muscle mass, function and quality among mild autonomous cortisol secretion (MACS) patients and non-functioning adrenal incidentaloma (NFAI) patients in comparison with the control group without adrenal mass.

Methods: 63 NFAI (49 female, 14 male) and 31 MACS (24 female, 7 male) patients were included in the study. As the control group, 44 patients (31 women, 13 men) who were known to have no radiological adrenal pathology on computed tomography or magnetic resonance imaging performed for other reasons were selected.

View Article and Find Full Text PDF

18F-Sodium Fluoride PET/CT as a Tool to Assess Enthesopathies in X-Linked Hypophosphatemia.

Calcif Tissue Int

January 2025

Endocrinology Department, School of Medicine, Pontificia Universidad Católica de Chile, Av. Diagonal Paraguay 262, Cuarto Piso, Santiago, Chile.

X-linked hypophosphatemia (XLH) is a rare metabolic disorder characterized by elevated FGF23 and chronic hypophosphatemia, leading to impaired skeletal mineralization and enthesopathies that are associated with pain, stiffness, and diminished quality of life. The natural history of enthesopathies in XLH remains poorly defined, partly due to absence of a sensitive quantitative tool for assessment and monitoring. This study investigates the utility of 18F-NaF PET/CT scans in characterizing enthesopathies in XLH subjects.

View Article and Find Full Text PDF

Correlations between spinopelvic parameters and health-related quality of life in degenerative lumbar scoliosis patients before and after long -level fusion surgery.

BMC Musculoskelet Disord

January 2025

Department of Orthopedics, Peking University Third Hospital, No. 49. North Garden Street, Hai Dian District, Beijing, 100191, People's Republic of China.

Background: For degenerative lumbar scoliosis (DLS), prior studies mainly focused on the preoperative relationship between spinopelvic parameters and health-related quality of life (HRQoL), lacking an exhaustive evaluation of the postoperative situation. Therefore, the postoperative parameters most closely bonded with clinical outcomes has not yet been well-defined in DLS patients. The objective of this study was to comprehensively assess the correlation between radiographic parameters and HRQoL before and after surgery, and to identified the most valuable spinopelvic parameters for postoperative curative effect.

View Article and Find Full Text PDF

Although the Transformer architecture has established itself as the industry standard for jobs involving natural language processing, it still has few uses in computer vision. In vision, attention is used in conjunction with convolutional networks or to replace individual convolutional network elements while preserving the overall network design. Differences between the two domains, such as significant variations in the scale of visual things and the higher granularity of pixels in images compared to words in the text, make it difficult to transfer Transformer from language to vision.

View Article and Find Full Text PDF

Functional magnetic resonance imaging (fMRI) has dramatically advanced non-invasive human brain mapping and decoding. Functional near-infrared spectroscopy (fNIRS) and high-density diffuse optical tomography (HD-DOT) non-invasively measure blood oxygen fluctuations related to brain activity, like fMRI, at the brain surface, using more-lightweight equipment that circumvents ergonomic and logistical limitations of fMRI. HD-DOT grids have smaller inter-optode spacing (~ 13 mm) than sparse fNIRS (~ 30 mm) and therefore provide higher image quality, with spatial resolution ~ 1/2 that of fMRI, when using the several source-detector distances (13-40 mm) afforded by the HD-DOT grid.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!