We evaluated the accuracy of a wearable sensor vest for real-time monitoring of physiological responses to treadmill exercise. Ten subjects in standard firefighter ensembles, treadmill exercising at 50% VO(2) max, had heart rate (HR), respiratory rate (RR), skin temperature (T(sk)), oxygen saturation (SaO(2)), tidal volume (V(T)), and minute ventilation (V(E)) recorded concurrently by a wearable plethysmographic sensor vest and standard laboratory physiological monitoring equipment for comparison. A high degree of correlation was noted for most of the measured variables [HR (r = 0.99), RR (r = 0.98), T(sk) (r = 0.98), V(E) (r = 0.88), and SaO(2) (r = 0.79)]. V(T) (r = 0.60) had a moderate correlation, although a paired differences analysis showed a mean paired difference of -0.03 L. This mean paired difference represents a 1.92% variation for V(T). Data from the wearable sensor vest is comparable to data captured from standard laboratory physiological monitoring equipment on subjects wearing standard firefighter ensembles while exercising at a moderate work rate. This study demonstrates the accuracy of the wearable sensor technology for these physiological parameters under these conditions and suggests that it could be useful for actual field studies of firefighters in traditional firefighting gear.

Download full-text PDF

Source
http://dx.doi.org/10.1080/15459620903455722DOI Listing

Publication Analysis

Top Keywords

sensor vest
16
physiological monitoring
12
firefighter ensembles
12
wearable sensor
12
wearable plethysmographic
8
plethysmographic sensor
8
accuracy wearable
8
standard firefighter
8
standard laboratory
8
laboratory physiological
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!