Introduction: We aimed to compare plasmakinetic prostate resection (PKRP), plasmakinetic prostate vaporization (PKVP) and transurethral prostate resection (TURP) procedures according to cautery artifacts in tissue specimens and to compare the coagulation depths of these procedures.

Materials And Methods: The prostate specimens of 10 patients that underwent open transvesical prostatectomy were used. TURP, PKVP and PKRP procedures were performed immediately. Artifactual pathological patterns that were identified in the specimens included: abnormal cellular orientation and spindling, artifactual cellular detachment from the underlying basement membrane, atypical cytological changes and stromal coagulative artifacts. The severity of cautery artifact was graded as absent, mild, moderate or severe according to the sum of points in each specimen.

Results: When the groups were compared according to moderate and severe artifacts, it was observed that PKRP caused more moderate artifacts and TURP caused more severe artifacts than the other two groups (p < 0.05). The depth of the tissue affected by coagulation was 1.52 +/- 1.29 mm with the TURP procedure. There were significant differences between TURP and the other two methods (p < 0.05).

Conclusions: Three transurethral procedures cause cautery artifacts of varying grades. The application of TURP seems to cause more severe artifacts and PKRP and PKVP procedures seem to result in a deeper coagulation zone in the residual prostatic tissue.

Download full-text PDF

Source
http://dx.doi.org/10.1159/000251188DOI Listing

Publication Analysis

Top Keywords

severe artifacts
12
prostatic tissue
8
plasmakinetic prostate
8
prostate resection
8
procedures cautery
8
cautery artifacts
8
moderate severe
8
artifacts
7
turp
6
vitro study
4

Similar Publications

Rationale And Objectives: The aim of this study was to compare the image quality of a deep learning (DL)-accelerated volumetric interpolated breath-hold examination (VIBE) sequence with a standard (ST) VIBE sequence in assessing the uterus.

Materials And Methods: Between April and December 2023, a total of 61 female patients (aged 41 ± 14 years) who were referred for an magnetic resonance imaging (MRI) of the pelvis were included in this prospective study, after providing informed consent. All examinations were performed with a 1.

View Article and Find Full Text PDF

Purpose: To develop and evaluate a physics-driven, saturation contrast-aware, deep-learning-based framework for motion artifact correction in CEST MRI.

Methods: A neural network was designed to correct motion artifacts directly from a Z-spectrum frequency (Ω) domain rather than an image spatial domain. Motion artifacts were simulated by modeling 3D rigid-body motion and readout-related motion during k-space sampling.

View Article and Find Full Text PDF

Photoacoustic imaging (PAI) is an emerging hybrid imaging technology that combines the advantages of optical and ultrasound imaging. Despite its excellent imaging capabilities, PAI still faces numerous challenges in clinical applications, particularly sparse spatial sampling and limited view detection. These limitations often result in severe streak artifacts and blurring when using standard methods to reconstruct images from incomplete data.

View Article and Find Full Text PDF

Accuracies of measuring the artifact index (AI), a quantitative artifact evaluation index in X-ray CT images, were investigated. The AI is calculated based not only on the standard deviation (SD) of the artifact area in the image, but also on the SD of noise components for considering the noise influence. However, conventional measurement methods may not follow this consideration, for example the non-uniformity of the noise distribution is not taken into account, resulting in reducing the accuracy of AI.

View Article and Find Full Text PDF

Artifact Reduction in Interventional Devices Using Virtual Monoenergetic Images and Iterative Metal Artifact Reduction on Photon-Counting Detector CT.

Invest Radiol

January 2025

From the Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Bonn, Germany (Y.C.L., N.M., P.A.K., A.I., T.D., J.A.L., D.K.); and Siemens Healthineers AG, Erlangen, Germany (S.F., V.H., B.S.).

Objectives: The aim of this study was to assess the impact of an iterative metal artifact reduction (iMAR) algorithm combined with virtual monoenergetic images (VMIs) for artifact reduction in photon-counting detector computed tomography (PCDCT) during interventions.

Materials And Methods: Using an abdominal phantom, we conducted evaluations on the efficacy of iMAR and VMIs for mitigating image artifacts during interventions on a PCDCT. Four different puncture devices were employed under 2 scan modes (QuantumSn at 100 kV, Quantumplus at 140 kV) to simulate various clinical scenarios.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!